Model DT23x-0600
Dual Two-Wire DC I/V Input Transmitters
Acromag, Inc. Tel: 248-295-0880
- 14 -
http://www.acromag.com
- 14 -
https://www.acromag.com
CONFIGURATION SOFTWARE
Quick Overview
–
Android Reconfiguration
This transmitter can be optionally setup & calibrated via the
Acromag Agility™ Config Tool. This software APP can be
downloaded free of charge from
. To connect to
this transmitter, a USB OTG (On-The-Go) cable (5028-565) and USB
A to Mini-B cable (4001-113) are required. This app is compatible
with Android devices using Ice Cream Sandwich (4.0) or later.
The initial Connection screen of the app is shown at left. Once a device is
connected, the main portion of the app will launch. Briefly, the APP screen is
divided into four tabs for this model. A short description of each tab follows.
Initial Connection Screen Set up
–
DEVICE SELECT (First Connect to Unit Here)
•
Select from connected transmitters by tapping the
[Select Device]
button
to bring up a list of attached devices. Select a device and tap
[Connect]
to
open the device.
•
To view wiring diagrams for this model, tap the
[Wiring Diagram]
button
and select the desired model. Swipe left or right to view a connection
drawing (note that no connection is required to just view drawings).
•
Android devices normally require user permission to access external
hardware. If
“No Device Permission” is indicated in the
Device List, select
this device and when prompted, grant permission to access the USB
device and tap
[OK]
.
Configuration Tab
–
(Setup your I/O)
•
Once connected, the app will automatically read your transmitter, its
mode, and display its current configuration.
•
Changing any option on this page will send the changes to the transmitter
instantly. The device status field at the bottom of the page will report if
the changes were sent successfully.
Calibration Tab
–
(Calibrate the Input and/or Output if Needed)
•
On screen instruction guides the set up to properly calibrate a transmitter
channel input or output. After completing instructions, tap the
[Calibrate]
button.
•
The device status at the bottom of the page will report if the calibration
was sent successfully.
Diagnostic Center Tab
–
(Verify Input operation)
•
Select the polling indicator by tapping the
[Indicator]
button.
•
Start polling by tapping the
[Start Polling]
button.
Utility Page
–
(Reboot or Restore Settings)
•
Tap the
[Gear]
in the Action bar to access the Utility Page.
•
You can tap the
[Restore/Reset Factory]
utility buttons to get out of
trouble if you ever misconfigure or improperly calibrate a transmitter.