OWL Data Loggers
23
Copyright © 2010 ACR Systems Inc. All Rights Reserved.
produces an output of approximately 260 microamps full scale for each range available on the
current probe. Readings from the current probe are scaled into amperage units through
equations (in TrendReader software) that match the current probe and its range setting.
1.5.2
Setup and Use
The procedure for setting up the OWL Electric Current Data Logger is as follows:
1.
Plug the current probe into the receptacle on the logger.
2.
Adjust the range switch on the current probe to suit your application. A good way to verify
a conductor's maximum range is to check the circuit breaker or fuse rating. Then set the
range on your current probe to suit. Since the probes do not give predictable readings
below 10% in any range, always choose the lowest range possible to encompass the full
range of data you want to record.
3.
Run TrendReader Software and choose the equation number that matches the probe
being used - Refer to Table 3-2. Then select the range corresponding to the logger's
switch setting. Note the software shows the highest possible reading which allows for a
small amount of overshoot. Therefore, if the switch setting is 5 on an A60FL/A65FL then
choose the 5.438 A range in software.
4.
Refer to the
WARNING: DO NOT CHANGE RANGE SWITCH WITHOUT ALSO SETTING LOGGER
Table 3-2 Current Probe Ranges
A60FL / A65FL
Equation 206
A70FL / A75FL
Equation 207
Range
Switch
Full Scale
Switch
Full Scale
0
5
5.438 A
10
10.88 A
1
25
27.19 A
50
54.38 A
2
100
108.8 A
250
271.9 A
3
250
271.9 A
500
543.8 A
Precautions
1.
DANGER: High voltage may be present!
Do not clamp a current probe around a
conductor without the logger connected to it. This avoids sparking and prevents the
connectors from prematurely deteriorating.
2.
The OWL Electric Current logger can only be used with current probes from ACR
Systems (altered Amprobe Instruments models A60FL, A70FL, A65FL and A75FL).
These probes provide low voltage DC current in proportion to the signal being measured.
They have a convenient snap connector added to protect against harsh environments.
DO NOT use any other probe unless approved by ACR in writing.
3.
The OWL Electric Current logger can record current in accordance with the ranges
Summary of Contents for OWL 100
Page 3: ...TrendReader is a Registered Trademark of ACR Systems Inc TRADEMARK NOTICES ...
Page 4: ......
Page 44: ...OWL Data Loggers 37 Copyright 2010 ACR Systems Inc All Rights Reserved ...
Page 45: ...OWL Data Logger 38 Copyright 2010 ACR Systems Inc All Rights Reserved ...
Page 46: ...OWL Data Loggers 39 Copyright 2010 ACR Systems Inc All Rights Reserved ...
Page 48: ...OWL Data Loggers 41 Copyright 2010 ACR Systems Inc All Rights Reserved ...
Page 49: ...OWL Data Logger 42 Copyright 2010 ACR Systems Inc All Rights Reserved ...
Page 50: ...OWL Data Loggers 43 Copyright 2010 ACR Systems Inc All Rights Reserved ...
Page 54: ......
Page 55: ......