APPENDIX I - ILLUSTRATED OPERATIONAL SEQUENCE
Procedure for Quick Start-Up
The following procedure can be used when the programs are going to be used as pre-set in the Omnicycle
®
Elite,
and when there is no need to modify the
parameters
.
When the cycle is in STAND-BY mode or is not
turned on, the Omnicycle
®
Elite can be activated by
pressing any button on the control panel. The screen
will light up. During a self test, which lasts about
five (5) seconds, the welcome screen is displayed,
followed by the system version screen. After the
self test, the unit will display the following start
screen:
The Omnicycle
®
Elite is now ready for use.
Pressing the START button at this point will start
the exercise session with the basic default settings,
with the pedals going at 10 rpm in assisted exercise
mode. By actively turning the pedals at speeds
greater than this, the cycle will change to active
exercise mode at any time.
By pressing the STOP button once during the
exercise session, the Omnicycle
®
Elite will pause.
It will resume the same session when the START button is pressed.
When the START button is pressed, and the cycle has started the exercise session, the following screen will be
displayed:
The DURATION will display the time remaining in
the started exercise session, and will count down to
zero, at which time the cycle will sound an audible
tone and stop. The exercise session activity report
screen will be displayed.