Serinus 60 User Manual 3.0
Page 54
Serial Delay
Some older communication systems require a delay before
the instrument responds to a serial command. The number of
milliseconds of delay required (0-1000). The default is 0,
meaning the instrument responds as quickly as possible to any
serial request.
Baudrate
Sets the baud rate for this serial port (
1200
,
2400
,
4800
,
9600
,
14400
,
19200
,
38400
or
115200
).
Protocol
Sets the protocol used for this serial port (
Advanced
,
ModBus
,
EC9800
or
Bayern-Hessen
). This must be set to
Advanced
for
Ecotech supplied software.
Endian
[Modbus Protocol]
Select
Little
or
Big
endian mode for ModBus protocol.
3.4.24
Analog Input Menu
Main Menu
Communications Menu
Analog Input Menu
The Serinus supports three analog inputs from the 25 pin I/O connector. Each input is a 0 to 5 volt
CAT 1 input that can be scaled and logged to the USB memory stick or accessed remotely as
parameters 3 to 5.
CAUTION
Exceeding these voltages can permanently damage the instrument and
void the warranty.
Input 1/2/3
The sections below are repeated for each analog input.
Multiplier
The input voltage will be multiplied by this number. For
example, if a sensor has a 0-5V output for a temperature of -
40 °C to 60 °C, the multiplier would be (60-(-40))/5 = 20.
Offset
This value will be added to the above calculation. Continuing
the example in the multiplier description, the offset should be
set to -40, so that a voltage of 0V will be recorded as -40°C.
Reading
The current reading from the input voltage, after the
multiplier and offset has been applied. This is the value that is
logged or reported as parameter 3 to 5 via USB or serial
requests.
3.4.25
Analog Output Menu
Main Menu
Communications Menu
Analog Output Menu
Output Mode
The analog output can be set to be either
Current
or
Voltage
.
Different fields will be displayed depending on which analog
output type is selected.
NO2
The name of the gas for each analog output. All of the fields
below are repeated for each gas.
Summary of Contents for Met One Instruments GAS-1060
Page 10: ...Serinus 60 User Manual 3 0 Page 10 This page is intentionally blank...
Page 18: ...Serinus 60 User Manual 3 0 Page 18 This page is intentionally blank...
Page 35: ...Installation Page 35 Figure 7 Switching the Battery On Off...
Page 36: ...Serinus 60 User Manual 3 0 Page 36 This page is intentionally blank...
Page 102: ...Serinus 60 User Manual 3 0 Page 102 This page is intentionally blank...
Page 134: ...Serinus 60 User Manual 3 0 Page 134 9 5 Plumbing Schematic PN D020060...
Page 135: ...Parts List Page 135 9 6 Block Wiring Schematic PN D020111...
Page 136: ...Serinus 60 User Manual 3 0 Page 136 9 7 Measurement Cell Side Section PN H015000...
Page 137: ...Parts List Page 137...
Page 139: ...Parts List Page 139 9 9 Valve Assembly Exploded View PN H010042...
Page 140: ...Serinus 60 User Manual 3 0 Page 140 9 10 High Pressure Valve Exploded View PN H050043...