The service partner shall issue a certificate on type and extend of annual inspection and repair. Inspection shall be marked 

with a label „next check“ on the rear panel of device. 

Opening  of  casing,  service  and  any  other  modifications  of  device  and  accessories  by  unauthorized  persons  releases 

manufacturer from any liability for the operating safety of the pump. 

Depending on respective country-specific regulations, a function and safety inspection must be performed by the user. 

Manufacturer recommends quarterly inspections, see chapter 12.

Each device is carefully tested with regard to function and safety by the manufacturer. Results are recorded in the test 
report which is attached to the device.

Information regarding schematics, component lists, descriptions and/ or calibration instructions can be provided by the 
manufacturer on demand.

15. QuarTerly InSpeCTIonS

15.1. CHeCkIng THe pneumaTIC ConTrol (manual ConTrol/Double FooTSWITCH)

For this test you will need: 

• Membrane suction-irrigation pump
• original disposable sterile filters (2x)
• Manometer with 1m tube and LL-connector female for pressure measurement (+)
• Manometer with 1m tube and LL-connector male vacuum measurement (-)

This test will be carried out without liquid.

Proceed as follows:

1. Connect the pump to main switch on device with power switch.
2. Activate the device by pressing button ON/OFF.
3. The device enters self test.
4. Connect manometer for pressure measurement to male LL-connector IRRIGATION and manometer for vacuum 
measurement to female LL-connector SUCTION.

A) Checking irrigation pressure:
5. Push button UP/DOWN IRRIGATION (8) to adjust preset pressure to maximum.
6. Start the pump.
7. Pressure increases to maximum and the pressure generator is switched off.
8. The pressure manometer shall display 750 mmHg +/- 10%.
9. The check of IRRIGATION pressure is complete. Press the button START/STOP to stop the pump

B) Checking suction vaccum:
10. Push button UP/DOWN SUCTION (11) to adjust preset vacuum to maximum.
11. Start the pump.
12. Vacuum increases to maximum and the vacuum generator is switched off.
13. The vacuum manometer shall display -525 mmHg+/-10%.
14. The check of SUCTION vacuum is complete. Press the button START/STOP to stop the pump.

In case of deviations check connections and repeat the test. If deviation is observed again refer to qualified technical 

service.

Summary of Contents for 16-2083

Page 1: ...SHAVER SYSTEM Ackermann www ackermannsurgical com Membrane Suction Irrigation Pump 16 2083 User Manual Membrane Suction Irrigation Pump ...

Page 2: ... of reusable tubing set 9 9 3 Illustration of tubing set 9 9 4 Connecting the tubing set 11 9 5 Removing the tubing set 11 9 6 Using the device in operating room 11 10 Footswitch 12 11 Functional buttons start up functional test and operation 13 11 1 Functional buttons 13 11 1 1 Power switch 13 11 1 2 On off 13 11 1 3 Start stop 13 11 1 4 Up down 13 11 2 Start up self test 14 11 3 Functional test ...

Page 3: ...with irrigation liquid separately 9 Display PRESSURE IRRIGATION available 10 Display VACUUM SUCTION 11 Buttons VACUUM UP DOWN SUCTION 12 LL connector female SUCTION connectors and overflow valve 13 Connector footswitch START STOP 14 Power switch 15 Fusebox 16 Mains connection 17 Equipotential bonding pin 18 Label next check 19 Model identification plate 20 Bottle holder 21 Bottle 22 SPIKE IRRIGATI...

Page 4: ...50 mm Hg Vacuum range 525 mm Hg Operating conditions Temperature 10 C to 40 C Humidity 30 to 70 Storage and transport conditions Temperature 40 C to 70 C Humidity 10 to 90 Case protection IPX1 Tubing sets Silicone autoclavable Dimensions without bottle holders 304 x 115 x 345 mm WxHxD with bottle holders ca 560 x 115 x 345 mm WxHxD Weight ca 9 kg Optional Footswitch Certification CE0483 3 Indented...

Page 5: ...ion irrigation tubing set autoclavable 16 2083 102 103 1 5 Main cable Euro 2 m 16 2083 104 1 6 Fuses 16 2083 105 2 7 Secretion container with overflow valve autoclavable 16 2083 101 1 8 Bottle holder 1 litre 2 screws for each 16 2083 106 2 9 Spike autoclavable 16 2083 109 1 10 Hydrophobic sterile filter sterile package disposable 16 2083 108 2 5 safety instructions disposal safety signs symbols Pl...

Page 6: ...operation of this device If the device is used in the vicinity of or stacked together with other devices the device or system must be monitored to ensure standard operation in accordance with these guidelines If necessary refer to manufacturer for further details WARNING The application of pump for small and closed body cavities arthroscopy and hysteroscopy is strictly forbidden because irrigation...

Page 7: ... spike 7 1 handling of spike The spike is needed to connect tubing set to the bottle with irrigation liquid Push spike completely into the bottle with irrigation liquid To ensure proper function venting hole must be placed inside the bottle but above liquid level Tip of spike must always be completely immersed in the liquid Connect tubing set IRRIGATION see chapter 9 4 Place bottle with inserted s...

Page 8: ...organic fluid and has a maximum capacity of 2 litres It is equipped with pressure lid mechanical overflow protection with integrated float and a volume scale Close the lid by pressing it onto the secretioncontainer Ensure that integrated float of overflow protection can move freely Connect tubing set SUCTION to the connectors VACUUM and PATIENT see chapter 9 4 Both connectors must not be mixed up ...

Page 9: ...ocess 9 2 reprocessing of reusable tubing set Read all instructions carefully before reprocessing the tubing set The entire user manual set consists of this product specific user manual delivered together with the pump and the user manual of the tubing set to be used for the pump Furthermore obey RKI guidelines and standards ISO17664 and ISO11607 Before reprocessing remove and dispose sterile filt...

Page 10: ...en device is switched on and self test is passed successfully Do not use tubing set without disposable sterile filters Start surgery with a new completely filled bottle of irrigation liquid prepare a spare bottle ATTENTION If the bottle is empty sterile air is transported to patient without any warning To remove a bottle first of all disconnect tubes from pump To prevent the instrument from suckin...

Page 11: ... SUCTION on pump Connect male LLconnector of the tube to disposable sterile filter Connect the other side of the tube to connector VACUUM on secretion container Step 4 Tube 4 red suction Connect one side of the tube to connector PATIENT on secretion container and the other side of the tube to suction connector on instrument 9 5 Removing the tubing set First of all disconnect tubes from pump to rel...

Page 12: ...ttle must be qualified for the overpressure of pump system 10 footswitch Optionally the pump can be operated by footswitch Footswitch has the same functionallity as button START STOP Choose a convenient position for the footswitch Connect the adapter of control line to connector 13 on front panel of pump Ensure correct and tight connection The control line must not be kinked or twisted Illustratio...

Page 13: ...FF The STANDBY LED 4 lights up For safety reasons button ON OFF is to press at least for 3 sec 11 1 3 start stop The push button START STOP 2 puts the pump into operation and stops it To start pump Press the button START STOP The START LED 1 lights up The pump is put into operation and the preset pressure vacuum are generated within the bottles To stop pump Press the button START STOP The STOP LED...

Page 14: ...prior to each operation Irrigation 1 Push button UP DOWN IRRIGATION 8 to adjust preset pressure to maximum 2 Start the pump 3 Pressure increases to maximum within ca 2 min and the pressure generator is switched off If preset pressure is not obtained check tubes and connections to be tight 4 Operate irrigation valve of instrument and fill the tubing set and the instrument completely with liquid unt...

Page 15: ...ed Device can not be operated Press button ON OFF to deactivate device and activate again after 15s Ensure that tubing set and footswitch are disconnected from device Should the error message still occur refer to qualified technical service In case of continuous beep and blinking LEDs during operation a device error was detected Pump stops immediately and can not be put into operation again Discon...

Page 16: ...s 2x Manometer with 1m tube and LL connector female for pressure measurement Manometer with 1m tube and LL connector male vacuum measurement This test will be carried out without liquid Proceed as follows 1 Connect the pump to main switch on device with power switch 2 Activate the device by pressing button ON OFF 3 The device enters self test 4 Connect manometer for pressure measurement to male LL...

Page 17: ...elf test 4 Connect footswitch to connector on frontpanel of the pump Ensure correct and tight connection The control line must not be kinked or twisted 5 Press footswitch to start the pump LED START lights up and pressure and vacuum generators are operating 6 Press footswitch to stop the pump LED STOP lights up and pressure and vacuum generators are switched off In case of deviations check setup a...

Page 18: ...tle holder 1 litre 2 screws 16 2083 107 Bottle holder 2 litres 2 screws 16 2083 108 Hydrophobic sterile filter sterile package disposable 16 2083 109 Spike autoclavable 16 2083 110 Footswitch 17 APPENDIX All product codes covered by these instructions are listed below Membrane suction irrigation pump 16 2083 18 CONTACT DETAILS Ackermann Instrumente GmbH Eisenbahnstrasse 65 67 78604 Rietheim Weilhe...

Page 19: ...Ackermann Page 19 20 SHAVER SYSTEM ...

Page 20: ...ietheim Weilheim Germany User_Manual 16 2083 Revision_06 21 Copyright by Ackermann Instrumente GmbH Specifications design and accessories in this user manual are subject to change without any notice or obligation on the part of the manufacturer Ackermann 0483 ...

Reviews: