SHAVER SYSTEM
If pressure and vacuum decrease below preset values they are switched on again.
If only one of both (IRRIGATION or SUCTION). Is needed you can switch off refering generator: push buttons UP/DOWN to
reduce preset pressure/ vacuum to zero.
5. Stop process by pressing the button START/STOP or the footswitch.
6. Deactivate device by pressing the button ON/OFF for 3 sec.
7. Switch off the device with power switch.
IMPORTANT!
Keep enough irrigation liquid, which is prewarmed to 40°C readily available. Use only intact, originally sealed
bottles with irrigation Liquid.
Operated bottles are pressurized! Danger of bursting!
ATTENTION!
If the irrigation bottle is empty, sterile air is transported to patient without any warning!
IMPORTANT!
Ensure to exchange secretion container in time! Integrated overflow valve shall protect device from being
contaminated by secretion Fluid. If in spite of all security precautions taken contamination occurs, switch off device
immediately and disconnect it from mains. Refer to qualified technical service.
12. WarnIng SIngnalS, FaIlure- & FaulT DeSCrIpTIon
• In case of continuous beep and blinking LEDs during selftest a device error was detected.
Device can not be operated.
* Press button ON/OFF to deactivate device and activate again after 15s. Ensure that tubing set and footswitch are
disconnected from device! Should the error message still occur refer to qualified technical service.
• In case of continuous beep and blinking LEDs during operation a device error was detected.
pump stops immediately and can not be put into operation again.
* Disconnect tubing set and footswitch from device. Press button ON/OFF to deactivate device and activate again after
15s. Should the error message still occur refer to qualified technical service.
13. CleanIng & maInTenanCe
To preserve the device and ensure its proper function please note as follows:
Cleaning:
Switch off the device with power switch. Disconnect device from mains. Remove power cable from mains connection at
the rear of device.
Use mild disinfecting solution to clean the exterior of device and footswitch. For fluid concentration follow instructions
from disinfectant manufacturer. Strictly prevent the ingress of liquid into the device and footswitch!
ATTENTION!
The device and the footswitch must not be sterilized!
14. ServICe
The unit must be checked by an authorized service partner once a year.
Repair, calibration and further services are to be carried out by authorized service only. Unauthorized services release
manufacturer from any liability for the operating safety of the pump.