
Smart Photoionization Detector Module PIDS3
User Manual
PIDS3 - User Manual (en) - 1.03.docx
Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com
Page 22 of 58
7
Installation hints
USAGE
The PIDS sensor is sensitive against pollution. Use dust and water filter to protect the gas
inlet of the module.
NOTICE
–
WATCH THE GAS FLOW
The module requires a uniform gas flow.
Do not enter the MEASURE state without appropriate gas flow.
USAGE
–
PLACE THE MODULES IN A THERMALLY ISOLATED ENVIRONMENT
The PID sensor is sensitive against fast gradients of temperature and humidity. To achieve the
best accuracy, it is recommended that the module is place in a thermal insulated
environment.
NOTICE
The module contains electronic components, which react sensitively to electrostatic
discharge (ESD). Work on or in the unit must be done only by qualified personal and in full
compliance with the appropriate instructions and pertinent regulations.