Smart Photoionization Detector Module PIDS3
User Manual
PIDS3 - User Manual (en) - 1.03.docx
Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com
Page 16 of 58
6.1
Installation
The device should be installed where gas leaks are expected. The installation position depends on the gas
density, either in the upper area of the room under the ceiling for gases lighter than air or close to the
ground for gases heavier than air. Also, consider how air movement may affect the ability of the device to
detect gas. The display on the front of the instrument must always be clearly visible; the view must not be
obstructed.
Before beginning the installation, check that the delivered components are complete and
correct referring to the shipping documents and the sticker on the shipment carton.
When preparing the assembly, make sure that the mounting arrangement is correct for the
device.