Acea NovoCyte Quick Start Manuals Download Page 1

NovoCyt

e

®

 and NovoCyte

®

 Quanteon

 Flow Cytometer

s

QUICK START & SAMPLE ACQUISITION GUIDE

STARTUP

STARTUP

RUN 

SAMPLES

QC TEST

1.

Check for adequate levels of the NovoFlow

®

, NovoRinse

®

, and

NovoClean™ in the fluidic containers. Ensure that each is at

least 1/2 full, and preferably empty waste container.

2.

Press the power button on the front panel of the 

flow

cytometer 

to

 

turn on the instrument. Launch the

NovoExpress

®

 software

 

on the computer. Startup is

automated and takes

 

approximately 6 minutes.  Check the

instrument status as

 

indicated by the LED indicator  on the

front panel:

Green 

= Normal.

The instrument is ready to use.

Orange 

= Warning.

Click on the status bar in the NovoExpress software to

 

review

the warning message.

Red 

= Error.

Click on the status bar to review the error message.

Refer to 

Section 5: Troubleshooting 

in the 

Flow

 

Cytometer

Operator’s Guide

.

QC TEST

1.

Add 1 drop of the NovoCyte QC particles to 

0.5

 mL of

 

dilution

buffer (0.

4

 mL PBS and 0.

1

 mL NovoRinse solution).

2.

On the main menu, click on 

Instrument 

tab → Under 

Operation

Section →  Click 

QC Test

 and follow the software prompts to

complete the test.

3.

If the QC Test results display Pass or Acceptable, the instru-

ment may be used to acquire data. If any QC Test results

display “Fail”, refer to 

Section 5: Troubleshooting 

in the

 

 Flow

Cytometer Operator’s Guide

.

SHUTDOWN

1.

Press  the power button on the front panel of the 

flow

cytometer

.

 

The instrument will automatically perform the

shutdown

 

cleaning process and then power off.

2.

When completed

 

the LED indicator will turn off.

FLUIDICS MANAGEMENT

1. The software will display a warning message instructing the

user to add more  reagents when the volumes of NovoFlow,

NovoClean, and NovoRinse are less than the set thresholds of

0.9 L, 150 mL, and 150 mL, respectively.

2. To add reagents:

1) Make sure the instrument is Off or in the Ready status.
2) Disconnect the tubing from the container.
3) Bring the container to the sink. Unscrew the cap and add

more reagent to the container.

4) Screw the cap back on, reconnect the quick coupler, and

return the container back to the fluidics station. Ensure

that the tubing is not twisted or kinked.

5) If the instrument was in the Ready status, click 

Instrument

tab

 → 

Under 

Fluidics Maintenance → 

Click 

Priming

 in the 

software to prime the system before running samples.

3. Similarly, the software will warn the user to empty the waste

container when the waste volume exceeds 2.8 L.

1) Follow the steps above to disconnect the waste container

from the instrument. Dispose of the contents according to

local safety guidelines.

2) Add 300 mL of bleach to the waste container and

reconnect the container to the instrument.

SETUP

SAMPLES AND PLOTS

VERIFY

CORRECT PARAMETERS AND 

ACQUISITION SETTINGS

Events/Volume/Flow Rate/Threshold

RUN

COMPENSATION 

CONTROLS

(if applicable)

© 201

8

 ACEA Biosciences, Inc. All rights reserved.