Manual Part # 800-1402-002 Rev K, 072418
Page 49 of 53
Pr39 through Pr44 apply to General Purpose E and 950 Series Readouts
(Firmware versions end
in ‘C’)
Pr39
– Non-Linear Compensation ·········································· [0 or 1]
This parameter Enables/Disables the
Default Value = 0
Pr40
– Non-Linear Compensation Interval ··················· [0.5 to 10.0 in.]
This parameter sets the distance interval used while performing non-linear
compensation calibration.
Default Value = 5.000 in. (127mm)
Pr41
– Temperature Compensation ····································· [0 or 1]
This parameter enables (disables) the
Default Value = 0
Pr42 & Pr43
– NO LONGER USED
Pr44
– Temperature Compensation Coefficient ··················· [10 to 999]
This parameter sets the temperature expansion coefficient used when
temperature compensation Pr41 = 1.
Default Value = 150
Range:
10 to 999 will yield values from 1.0 m/m °K x 10
-6
to 99.9 m/m °K x 10
-6.
Note:
Coefficients MUST be in m/m °K.
Pr 39 Setting
Non-Linear Compensation
0
Off
1
On
Pr 41 Setting
Temperature Compensation
0
Off
1
On