Pr5
– Movement to Wake Readout ···························· [0.10 to 10mm]
This parameter sets the amount of encoder, (or system), movement required to
automatically wake up the Readout when it is turned off or in sleep mode.
Default Value = 0.10mm
Pr6
– Fractional Rounding -------------------------------------------- [0 or 1]
This parameter is used to control rounding in Fractional display mode. When
set to 0, rounding down to the least full 1/64th is used. When set to 1, rounding
to the
nearest
1/64th is used. (i.e. 1/64th + .0078" or greater rounds
UP
to next
1/64th.)
Default Value = 0
Pr11
– Readout Measurement Units ······································· [0 to 6]
This parameter controls the type of measuring units the Readout displays when
the
UNITS
key is pressed. The table below illustrates the possible combinations
of measuring units that will be displayed by changing this parameter.
Default Value = 0
Pr12
– Readout Auto-Off Time ············································· [0 to 240]
This parameter changes the amount of time before the display turns off to
conserve battery power. The value is the number of minutes of idle operation
(no movement or key presses) before the display turns off. When this
parameter is enabled (
0
), pressing the
ON
/
OFF
key or movement of the
encoder or system will wake up the Readout.
A value of ‘
0
’ disables the Auto Off feature (the Readout is always on).
Default Value = 15 (minutes)
Pr13
– Linear Compensation ································ [0.00001 to 9.99999]
This parameter invokes a linear multiplier correction in the Readout that is
applied to the actual measurement prior to any offsets. This adjusted
measurement is then displayed on the Readout.
Default Value = 1.00000
Pr14
– ProScale Compatibility ················································ [0 or 1]
This parameter sets the Readout to operate with the correct technology.
This parameter is preset at the factory if the Readout is purchased as part of a
complete System. See also
Default Value = 1
Pr 11 Setting
Displayable Units
0
All inch units and millimeters
1
Millimeters only
2
Decimal inches and millimeters
3
Decimal inches and centimeters
4
All inch units and centimeters
5
Centimeters only
6
Decimal inches only