During the calibration process outlined in Step 7, the user may wish to verify
the correction point that is to be programmed next. This can be accomplished
by
Momentarily
pressing the
F3
key. The LCD will display
CP X
, where
X
is
the
next
correction point number to be programmed. This feature is helpful
after programming many correction points since it can be easy to lose track of
the number of standards programmed.
NOTE:
If an error is made during the calibration process where an incorrect
table entry is recorded, the entire process must be repeated.
Operating in Compensated mode.
After a successful correction table has been programmed, the Readout will
operate normally and the measuring system will have a higher level of
accuracy.
NOTE:
If the encoder is positioned outside the calibrated (corrected) range of
operation, the LCD will display
No Co
alternately with the current position.
This is normal and indicates that the system is operating outside the calibrated
measuring range and has a reduced measuring accuracy.
If you choose to turn off Non-Linear Compensation (
Pr39=0,
) the compensation
table will be saved in memory. Thereafter, changing
Pr39
to
1
will re-enable the
previously stored compensation table.
NOTE
the Readout to factory default settings does not erase the
compensation table.