Circuit Board Jumpers
JP1
FACTORY USE ONLY
JP2
FACTORY USE ONLY
Position A ONLY
JP3
Programming Lock-out
Default = Position A
Readout programming can be enabled or disabled though the use of this
circuit board jumper.
Programming is enabled when the shorting jumper is installed in position
A
.
To disable Readout Programming, install jumper on position B.
When programming is disabled, user cannot access the programming
functions via the
front panel readout keys as described in the
Section 4:
PROGRAMMING PARAMETERS
.
This provides a method of configuring the ProPanel
with specific parameters, and preventing unauthorized or accidental changes.
JP4
Power Selection
Default = Position B
The ProPanel standard Readout operates on either on internal battery power
(Recommended) or external 12-24VDC. When this jumper is installed in the
B
position, the Readout is powered from the internal battery. When this jumper
is installed in position A, the Readout must be powered by 12-24VDC DC.
JP5
Input/Output Selection
Default = Position A
The ProPanel readout has connections used for External Power or an
Input/Output connection. This jumper is used to configure the connections as
Inputs or Outputs. In position
A
, the connections are configured for OUTPUT.
In the B position, the connections are configured as INPUTS