62
08424705001 (01) 2017-04 USA • Accu-Chek Inform II Operator’s Manual Version 6
Results screen
The hourglass icon indicates the test is running. When
the test is completed and the result is ready, the meter
beeps again.
When the result is displayed, a message or warning may
also appear (depending on system configuration) notify-
ing you if the result exceeds the specified limit values.
Additionally, a
red flashing arrow
next to the test result
indicates that a result is out of range.
: the value is
above
the normal/critical upper range.
: the value is
below
the normal/critical lower range.
These limit values define ranges that can either be con-
figured individually by the system administrator in line
with facility guidelines, or are the (technical) limits of the
system. The characteristics of these ranges are explained
on the following page.
The results screen contains a button that changes its
name based on the result (
Range
or
Out of ... Range
).
Touch this button to display the configured limit values.
Patient Test
Patient 123456789
Date 03/17/17 12:48 pm
150
L22
12:48 pm
03/17/17
Range
I-502: Normal Range:
Critical Range:
Reportable Range:
70-200 mg/dL
40-301 mg/dL
11-600 mg/dL
Patient Test
Patient 123456789
Strip Lot 545794
12:48 pm
03/17/17
DOB: 12/05/70
Patient Test
Range
150
Patient 123456789
Date 03/17/17 12:48 pm
mg/dL
12:48 pm
03/17/17