COVID-19 Treatment Bed
INSTRUCTION MANUAL (CV1)
Welcome
The Covid-19 treatment bed (CV1) is a nursing bed classified as a Class 1 Medical
Device in accordance with the Medical Device Regulation 2017/745.
Intended Use
The Covid-19 bed is intended for an adult population and can be used in all fixed and
temporary hospital and nursing environments.
WARNINGS
•
Always keep this manual available for future reference.
•
Only use compatible mattresses with this bed. Refer to specification for
dimensions.
•
A patient risk assessment must be carried out before both the bed
frame and / or the side rails are used.
•
Always fully decontaminate the bed in between users.
•
Always lock the folding arm of the legs in place by using the locking
ring to slide over the hinge at each end of the bed.
•
Care should be used when using the CPR Handle, there is a risk of
entrapment when the backrest falls to flat position.
Notice to user
If a serious incident occurs in relation to this medical device, affecting the user or the patient,
then the user or patient should report the serious incident to the medical device manufacturer
(or distributor) and, in the European Union, the user should also report the serious incident to
the Competent Authority in the member state where they are located.
IV Pole (Option: IVP-0-CV1-000)
The IV pole should be fitted to the bed frame at the head end of the bed:
Height Extension (Option: KITHA-0-CV1-000)
Maintenance, Guarantee & Disposal
Regularly check: side supports are fitted securely; legs are locked in position. The
warranty period for the Covid-19 bed is 2 years. In the event of the disposal of the
Covid-19 bed, end-of-life parts must be disposed of in accordance with current
environmental regulations.
Specification & Decontamination
Description
Value
Bed frame fits mattress dimensions
900 x 2000 mm
Height
450 mm platform 150 mm mattress
Safe working load /
Max. patient weight
185 kg /
150 kg
Backrest positions
Pre-set positions: 30/45/60°
Cleaning and decontamination
Use hospital-grade disinfection protocols
Symbols:
Design Policy and Copyright
® And ™ are trademarks belonging to Accora Ltd unless otherwise stated. As our policy is
one of continuous improvement, we reserve the right to modify designs without prior notice.
IFU-CV1-001EN Rev 01
Accora Limited, Barrington Road, Orwell,
Cambridge SG8 5QP, United Kingdom
Tel: 01223 206100
Website: www.accora.uk.com
Place the IV pole
brackets over the
main bed frame
Insert screws
(x2) from
inside and
tighten
1
x2
2 bolts & wing
nuts each end
Tighten the
wing nuts
2
3
IV Pole (Option:
IVP-0-CV1-000
)
Headboard
CPR Handle
Manual Backrest
Side Rails set (3 positions):
Head End, Middle, Foot End
(Option:
SDR-0-CV1-000
)
Footboard
Folding legs
(locked)
Bed frame
COVID-19 Treatment Bed (CV1)
Part No: NSB-0-CV1-000
1
4
Refer to Instruction Manual
Physical description of Adult