Maintenance
& Troubleshooting
.Replacement procedures
Pentra
80 - User Manual - RAB108EA
7–15
Place the sample tubing back onto the top of the sample probe.
Re-insert the top retaining part of the sample probe back into its retaining slot.
Lower the locking arm to secure the probe.
Tighten the Rinse block retaining screws for a good seal on the sample probe.
Power on the instrument and verify that there are no fluid leaks during startup. Re-install the
right-hand side panel.
2.4. Rinse chamber filter cleaning
1- This filter has to be cleaned once a month as follows:
2- Dismantle the Right hand side panel as described in
page 7-18, in order to access to dilution chambers.
Fig. 7–16 Rinse chamber filter location
3- Locate the filter below the rinse chamber.
4- Remove the filter, disconnecting tubings, and open it unscrewing both parts.
Fig. 7–17 Dismantling the filter
5- Rinse both parts of the filter under tap water.
6- Leave the filter dry, and re-install reverse order.
Summary of Contents for Pentra 80 RAB108EA
Page 1: ...User Manual P n RAB108EA ABX Diagnostics B P 7290 34184 MONTPELLIER Cedex 4 FRANCE...
Page 12: ...10 Section9 Glossary 1 Glossary 9 2 2 Index 9 4 3 9 11...
Page 87: ...Quality Assurance Calibration Pentra 80 User Manual RAB108EA 3 31 Fig 3 29 Delete options...
Page 162: ...Pentra 80 5 10 Pentra 80 User Manual RAB108EA Fig 5 6 Units tab...
Page 194: ...Pentra 80 5 42 Pentra 80 User Manual RAB108EA...
Page 214: ...Pentra 80 6 20 Pentra 80 User Manual RAB108EA...
Page 264: ...Pentra 80 7 50 Pentra 80 User Manual RAB108EA...
Page 284: ......
Page 286: ......
Page 288: ......
Page 290: ......
Page 292: ......
Page 294: ......
Page 296: ......
Page 306: ...Pentra 80 9 10 Pentra 80 User Manual RAB108EA Z Zoom Chap 4 28...