1.6. Sample identification on Rack/position
Setting: «
Rack / Position
, page 5-7, i.e. that the
identifier is given by the sample position on the rack.
When a Barcode label is not used on a sample tube, the order must contain the sample tube
position in the rack. The «Sample ID» field is now available for manual entry, but it is not
mandatory to do so.
◆
If the sample tube position in the rack was indicated, the final results will contain sample
identification.
◆
If not, the instrument assigns an auto numbering («AUTO_SID_xxx) to ensure that the final
results for this sample contains a single identification.
1.6.1. Setting: Rack/position / Manual match = OFF
The P80 tab: «
Settings/Soft parameters/General
tab/barcode option
» is set as shown: