Cecomp
®
Intrinsically Safe Digital Pressure Gauges, with Min/Max Options
DPG2000B D4 and M0, M1 Series
O
±0.25% Test Gauge Accuracy
O
316L Stainless Steel Sensor
O
All Metal Housing
Agency Approval
Factory Mutual Approved Intrinsically Safe for Hazardous Locations
USA & Canada
IS Class I, Division 1, Groups A, B, C, D
T3C Ta = –40°C to 82°C; T4 Ta = –40°C to 66°C
CL I Zone 0 AEx/Ex ia IIC
T3 Ta = –40°C to 82°C; T4 Ta = –40°C to 66°C
Ranges and Resolution
See table below. Select range and default engineering unit.
Units may be changed to any listed under the same sensor range.
Resolution is fixed and limited to available display digits.
Display
3 readings per second nominal display update rate
4 digit LCD, 0.5" H and 5 character 0.25" H alphanumeric
BL
: Red LED backlight. Keypress activates backlighting for 1
minute if ambient light is insufficient.
Accuracy
Accuracy includes linearity, hysteresis, repeatability
Standard accuracy: ±0.25% of full scale ±1 least significant digit
HA
accuracy option: ±0.1% FS ±1 LSD, see ranges for availability
Sensor hysteresis: ±0.015% FS, included in accuracy
Sensor repeatability: ±0.01% FS, included in accuracy
Auto Shutoff
5, 10 , 30 minutes, or on/off. User selectable 1 minute to 8 hrs
Memory Options
Min/max can be user configured to be individually enabled or dis-
abled, readings saved or cleared at power off
Std: Min/max enabled
M0
: Min/max turned off in user setup
M1
: Peak reading only gauge for tire pressure
Calibration
Non-interactive zero, span, and linearity, ±10% of range Internal
setup/calibration buttons, internal lockout switch.
Controls and Functions
Front button turns gauge on or off, zeros gauge reference gauges,
and cycles through min/max functions (if enabled)
Internal buttons for engineering unit selection, auto shutoff time,
min/max setup, calibration
Internal lockout switch to disable setup and calibration
Batteries
Two 1.5 V AAA Panasonic LR03 alkaline cells
B
:
Approx. 1000 hours
BL
: Approx. 150 to 1000 hours depending on backlight usage
Low battery symbol on display when batteries need replacement
Weight
9 ounces (approx.), shipping wt. 1 pound (approx.)
Housing Materials and Circuit Board Protection
Epoxy powder coated aluminum case, rear cover, and bezel. Front
and rear rubber gaskets, polycarbonate label.
Stainless steel stiffener plate to reinforce sensor area.
Conformal coating on circuit boards for moisture resistance.
Connection and Material
1/4" NPT male fitting. All wetted parts are 316L stainless steel.
Overpressure, Burst, Vacuum
2 X pressure range for 3 psi to 2000 psi sensors
5000 psig for ranges using 3000 psig sensor
7500 psig for ranges using 5000 psig sensor
Over-range display 112.5% FS:
I – – –
or
I –.–.–.–
Under-range display (non-vacuum sensors):
–Err
Vacuum service:
15 psia, ±15 psig, 15 psig, 30 psia,
100 psig, 100 psia, 200 psig sensors
Burst: 4 X sensor pressure rating or 10,000 psi, whichever is less
Environmental
Storage temperature:
–40 to 203°F (–40 to 95°C)
Operating temperature:
–4 to 185°F (–20 to 85°C)
Compensated temperature: 32 to 158°F (0 to 70°C)
Sensor Ranges and Engineering Units
* -HA option not available
Range codes are rounded off
3 psig
*
Res
15 psig vac
*
Res
30 psia
Res –15V100psig
*
Res
300 psig
Res
3PSIG
.001
100KPAVAC
.1
2KGCMA
.001
100PSIVAC
.1
300PSIG
.1
6INHGG
.001
0.1MPAVAC
.0001
2ATMA
.001
100PSICPD
.1
610INHGG
.1
85INH2OG
.1
1BARVAC
.001
30 psig
Res
200INHGVAC
.1
4800ZING
1
50ZING
.01
1KGCMVAC
.001
30PSIG
.01
2770INH2OVAC
1
700FTH2O
.1
210GCMG
.1
1ATMVAC
.001
60INHGG
.01
1600ZINVAC
1
2000KPAG
1
150MMHGG
.1
15 psig
Res
850INH2OG
1
5200MMHGVAC
1
2MPAG
.001
150TORRG
.1
15PSIG
.01
480ZING
.1
5200TORRVAC
1
20BARG
.01
200MBARG
.1
30INHGG
.01
2100GCMG
1
700KPAVAC
1
20KGCMG
.01
200CMH2OG
.1
400INH2OG
.1
1600MMHGG
1
0.7MPAVAC
.001
20ATMG
.01
2000MMH2OG
1
240ZING
.1
1600TORRG
1
7BARVAC
.01
500 psig
Res
7FTH2O
.001
1000GCMG
1
2000MBARG
1
7KGCMVAC
.01
500PSIG
.1
20KPAG
.01
760MMHGG
.1
2100CMH2OG
1
7ATMVAC
.01
1020INHGG
1
5 psig
*
Res
760TORRG
.1
70FTH2O
.01
100 psig
Res
1150FTH2O
1
5PSIG
.001
1000MBARG
1
200KPAG
.1
100PSIG
.1
3500KPAG
1
10INHGG
.01
1000CMH2OG
1
0.2MPAG
.0001
200INHGG
.1
3.5MPAG
.001
140INH2OG
.1
35FTH2O
.01
2BARG
.001
2770INH2OG
1
35BARG
.01
80ZING
.1
100KPAG
.1
2KGCMG
.001
1600ZING
1
35KGCMG
.01
350GCMG
.1
0.1MPAG
.0001
2ATMG
.001
7000GCMG
1
35ATMG
.01
260MMHGG
.1
1BARG
.001
60 psig
Res
5200MMHGG
1
1000 psig
Res
260TORRG
.1
1KGCMG
.001
60PSIG
.01
5200TORRG
1
1000PSIG
1
350MBARG
.1
1ATMG
.001
120INHGG
.1
7000MBARG
1
2040INHGG
1
350CMH2OG
.1
±15 psig
*
Res
1660INH2OG
1
7000CMH2OG
1
2300FTH2O
1
3500MMH2OG
1
±15PSIVAC
.01
960ZING
1
230FTH2O
.1
7000KPAG
1
12FTH2O
.01
15PSICPD
.01
4200GCMG
1
700KPAG
.1
7MPAG
.001
35KPAG
.01
±30INHGVAC
.01
3100MMHGG
1
0.7MPAG
.0001
70BARG
.01
15 psia
Res
±400INH2OVAC
1
3100TORRG
1
7BARG
.001
70KGCMG
.01
15PSIA
.01
±240ZINVAC
.1
4100MBARG
1
7KGCMG
.001
70ATMG
.01
30INHGA
.01
±1000GCMVAC
1
4200CMH2OG
1
7ATMG
.001
2000 psig
Res
400INH2OA
.1
±760MMHGVAC
1
140FTH2O
.1
–15V200 psig
*
Res
2000PSIG
1
240ZINA
.1
±760TORRVAC
1
400KPAG
.1
200PSIVAC
.1
4070INHGG
1
1000GCMA
1
±1000MBARVAC
1
0.4MPAG
.0001
200PSICPD
.1
4600FTH2O
1
760MMHGA
.1
±1000CMH2OVAC
1
4BARG
.001
400INHGVAC
.1
14MPAG
.01
760TORRA
.1
±100KPAVAC
.1
4KGCMG
.001
5500INH2OVAC
1
140BARG
.1
1000MBARA
1
±0.1MPAVAC
.0001
4ATMG
.001
3200ZINVAC
1
140KGCMG
.1
1000CMH2OA
1
±1BARVAC
.001
100 psia
Res
1400KPAVAC
1
140ATMG
.1
100KPAA
.1
±1KGCMVAC
.001
100PSIA
.1
1.4MPAVAC
.001
3000 psig
Res
0.1MPAA
.0001
±1ATMVAC
.001
200INHGA
.1
14BARVAC
.01
3000PSIG
1
1BARA
.001
30 psia
Res
2770INH2OA
1
14KGCMVAC
.01
6100INHGG
1
1KGCMA
.001
30PSIA
.01
1600ZINA
1
14ATMVAC
.01
6900FTH2O
1
1ATMA
.001
60INHGA
.01
7000GCMA
1
200 psig
Res
20MPAG
.01
15 psig vac
*
Res
850INH2OA
1
5200MMHGA
1
200PSIG
.1
200BARG
.1
15PSIVAC
.01
480ZINA
.1
5200TORRA
1
400INHGG
.1
200KGCMG
.1
30INHGVAC
.01
2100GCMA
1
7000MBARA
1
5500INH2OG
1
200ATMG
.1
400INH2OVAC
.1
1600MMHGA
1
7000CMH2OA
1
3200ZING
1
5000 psig
Res
240ZINVAC
.1
1600TORRA
1
700KPAA
.1
480FTH2O
.1
5000PSIG
1
1000GCMVAC
1
2000MBARA
1
0.7MPAA
.0001
1400KPAG
1
35MPAG
.01
760MMHGVAC
.1
2100CMH2OA
1
7BARA
.001
1.4MPAG
.001
350BARG
.1
760TORRVAC
.1
200KPAA
.1
7KGCMA
.001
14BARG
.01
350KGCMG
.1
1000MBARVAC
1
0.2MPAA
.0001
7ATMA
.001
14KGCMG
.01
340ATMG
.1
1000CMH2OVAC
1
2BARA
.001
14ATMG
.01
How to Specify
Display Memory
DPG2000B
range -
D4
-time-options
-
Min/max
DPG2000BBL
range -
D4
-time options
backlit
Min/max
DPG2000B
range -
D4
-
M0
-time-options
-
Min/max off
DPG2000BBL
range -
D4
-
M0
-time-options
backlit
Min/max off
DPG2000B
range -
D4
-
M1
-time-options
-
Peak read
DPG2000BBL
range -
D4
-
M1
-time-options
backlit
Peak read
G = gauge reference pressure
CPD = compound; inHg vacuum, psi pressure
VAC = gauge reference vacuum
A = absolute reference
psi = PSI
inHg = INHG
oz/in
2
= ZIN
inH
2
O = INH2O
ftH
2
O = FTH2O
mmHg = MMHG
torr = TORR
mmH
2
O = MMH2O
kg/cm
2
= KGCM
g/cm
2
= GCM
kPa = KPA
MPa = MPA
mbar = MBAR
bar = BAR
cmH
2
O = CMH2O
atm = ATM
Range
—See table at left. Select a range code for default units.
Please specify if vacuum gauge requires a minus sign.
Time
—auto shutoff time
5
5 minutes. Default if not specified.
10
10 minutes
30
30 minutes
ON
No auto shutoff. On/off via front button.
xH
Custom shutoff time where x = up to 8 hours
Options
—add to end of model number. See price list for details.
HA
High accuracy, ±0.1% FS ±1 LSD. See range table.
PM
Panel mount, 4.1" x 4.1"
TP
Top port, gauge port on top of case
CD
Calibration data; 5 test points and date
NC
NIST traceability documentation, 5 points and date
H H H H H H
H H H H H
H H H H H H
H H H H H
H H H H H H
H H H H H
H H H H H H
H H H H H
H H H H H H
Made in USA
Quick Link
cecomp.com/is
Accessories
—order separately
CON14SS
Quick connector to install or remove gauge with-
out tools. 304 stainless steel, urethane seal.
SCR14SS
Filter screen fitting keeps debris out of gauge
sensor. For food vacuum packaging applications.
303SS body, 100 micron 304SS screen.
RB
High visibility orange rubber boot protects gauge
for portable applications. Not available with NEMA
4X models.
TP
Top gauge port. Primarily used with tire pressure
applications. Not available with NEMA 4X models.
© 08-18
cecomp.com
1220 American Way Libertyville, IL 60048
800-942-0315
CECOMP
Div. of
A