Absolute Process Instruments APD 41391 Quick Start Manual Download Page 1

How to Order

Factory Ranged and Configured

Options

Model

Power

Input

Outputs

Alarm Code

Description

APD 41391

85-265 VAC, 60-300 VDC

Specify 

T/C type,  

temperature 

range  in  

°F or °C

Ch. 1 

 

specify  

VDC or mADC 

range

Custom  

Push-to-Test 

___% span

Ch. 2  

specify 

T/C type

none

-

B
N

RA

HP
HT

U

Upscale burnout (standard)
Downscale burnout
No burnout, last valid value

Reverse-acting alarm

Latching relay, power-off reset
Latching relay, push button reset

Conformal coating for moisture 
resistance

APD 41391 D

9-30 VDC, 10-32 VAC

APD 41391 H

85-265 VAC, 60-300 VDC

HI alarm 
(failsafe)

APD 41391 DH

9-30 VDC, 10-32 VAC

APD 41391 L

85-265 VAC, 60-300 VDC

LO alarm

APD 41391 DL

9-30 VDC, 10-32 VAC

Thermocouple Input to DC & Thermocouple Dual Output Splitter/Converter, Factory Ranged 

APD 41391

1 Input:

 

Any Thermocouple

2 Outputs:

  Process Voltage or mA + Any Thermocouple, Alarm Relay Options

 

O

Split or Convert a T/C to DC Output & Any T/C Output

 

O

Add Additional Isolated Output to Existing T/C

 

O

Zero and Span Output Calibration

 

O

Input LoopTracker

®

 LED

 

O

Sink/Source for DC Output

 

Q

Convert/Isolate Thermocouples for PLC Input, 
Control and/or Validation

 

Q

Interface Thermocouples with Panel Meters, PLCs, 
Recorders, Data Acq., DCS, & SCADA Systems

Applications

Thermocouple Input, Factory Set

T/C types: 

J, K, T, E, M, N, P, R, S, B, C, D, G

Temp. range: Full ANSI range, or specify °F or °C range 
Linearity: 

±0.1°C and 0.001°C resolution  

Linearization: Polynomial, 1°C segments for types M and P
T/C CJC:  

Automatic

T/C current:  Less than 10 µA, including burnout sense
T/C burnout:  Upscale (standard), 

B

: downscale, 

N

: none

Custom: 

Provide T/C millivolt data, °F or °C range

Status LEDs

LoopTracker: 

Variable brightness green for input level 

Yellow LED: 

Output 1 Push-to-Test status, error status 

Red/green LED:  Alarm state (with alarm option only)

DC Output, Channel 1, Factory Set

Linearized to temperature 
Voltage: 

0-1 V to 0-10 V (10 mA max.), ±1 V to ±10 V

Current: 

0-1 mA to 0-20 mA

 

20 V compliance, 1000 

 at 20 mA 

 

Sinking or sourcing output 

Loop power:  20 VDC nom., regulated, 25 mADC 
 

<10 mV

RMS

 max. ripple

Thermocouple Output, Channel 2, Factory Set

T/C types: 

J, K, T, E, M, N, P, R, S, B, C, D, G

 

Output limited by smallest ANSI T/C range 

Linearization: Polynomials, 1°C segments for types M and P 
T/C CJC:  

Automatic, can be factory disabled

Output Calibration

Zero and span potentiometers for each output, ±15% range

Output Push-to-Test (DC Output 1)

Front push button and terminals enables/disables test level 
output. Set to 75% of output span. Specify custom setting.

Output Resolution

16 bit

Output Ripple and Noise

Less than ±0.2% of span

Optional Alarm Relay

Single setpoint dual DPST contact sets, factory configured 
1 Form A (NO) and 1 Form B (NC) contact sets (4 terminals) 
May be field wired for Form C operation
8 A max @ 240 VAC 
resistive load 
Use external contact 
protection (RC snub-
ber) for inductive loads
Red/Green bi-color 
LED indicates alarm 
status 
One set point, 12 
turn potentiometer, 
0-100% of span 
One reset point, 12 turn potentiometer, 0-100% of span 
Default: HI alarm, non-latching, normal acting (failsafe)
Relay Test button toggles relay to opposite state or resets 
relay with HT option

Ambient Temperature Range and Stability

–10°C to +60°C operating ambient 
Better than ±0.02% of span per °C stability

Response Time

500 milliseconds minimum

Isolation

Full 4-way galvanic: input, output 1, output 2, 
power, 1200 V

RMS

 min. 

600 VACp or 600 VDC common mode protection 
75 dB minimum common mode rejection 
Simultaneous 50 Hz and 60 Hz rejection

Housing and Connectors

IP 40, requires vertical installation on a 35 mm DIN 
rail inside a panel or enclosure. Allow room for  
air flow. For use in Pollution Degree 2 Environment. 
Eight 4-terminal removable connectors,  
14 AWG max. wire size

Power

85-265 VAC, 50/60 Hz or 60-300 VDC, 3 W maximum 
D versions: 9-30 VDC or 10-32 VAC 50/60 Hz, 3 W maximum

Dimensions

1.78" W x 4.62" H x 4.81" D  
(45 mm W x 117 mm H x 122 mm D)

Quick Link: 

api-usa.com/tc

© 02-20

api-usa.com

1220 American Way Libertyville, IL 60048 
Phone: 

800-942-0315

 

Fax: 800-949-7502

A

A

BSOLUTE

BSOLUTE

 P

 P

ROCESS

ROCESS

 I

 I

NSTRUMENTS

NSTRUMENTS

Free Factory 

I/O Setup!

 H H H H H H

  H H H H H

 H H H H H H

  H H H H H

 H H H H H H

  H H H H H

 H H H H H H

  H H H H H

 H H H H H H

Made in USA

Applications Link 

api-usa.com/apps

Pb

Lead 

Free

Sink or Source  

Output for  

mADC Channel

Output Test for  

DC Channel

Universal Power

Optional Alarm 

Relay

Input 

LoopTracker 

LED

Zero and 

Span for 

Each Output

Status LED

Note: Temperature range is limited by the smallest ANSI 
range of the two T/C types selected.

Switching Current (A)

Switching Voltage (V)

3  5  10 

30 50 100  300 500

8

5

3

1

0.5
0.3

0.1

AC inductive load 

(cos 

f

 = 0.4)

AC resistive 

load

DC inductive load 

(L/R = 7 ms)

DC resistive 

load

Removable Plugs

See Wiring 

Diagrams on 

Next Page

1     2     3     4

5     6     7    8

9   10   11   12

13   14   15   16

25   26   27   28

29   30   31   32

21   22   23   24

17   18   19   20

Description

The APD 41391 accepts a thermocouple temperature input and 
provides one DC process output and one thermocouple output. 
This makes it useful to match up the available thermocouple 
type with an instrument input type, for data recording, or when 
isolating the input.
The thermocouple input type and temperature range are fac-
tory set. The DC output type and range, and the thermocouple 
output type and range are factory set. The DC output and the 
thermocouple output correspond to the input temperature 
range specified. Mixed T/C types are limited by the ANSI 
temperature range. Use appropriate thermocouple extension 
wire as needed.
The input is sampled, CJC compensated, digitally converted to 
a linearized temperature signal, and then passed to the output 
stages where it is converted to a DC output and a simulated 
thermocouple output with CJC compensation. Contact factory 
if the CJC compensation is to be disabled.

Full 4-way galvanic isolation (input 1, output 1, output 2, 
power) make this module useful for ground loop elimination, 
common mode signal rejection, and noise pickup reduction. 

LoopTracker 

An API exclusive feature includes a green LoopTracker LED that 
varies in intensity with changes in the input signal and a red 
LoopTracker LED that varies in intensity with changes in the 
channel 1 DC output signal. It provides a quick visual picture 
of your process at all times and can greatly aid in saving time 
during initial startup and troubleshooting. 

Output Push-to-Test 

An API exclusive feature includes an output test switch to pro-
vide a fixed output for the DC output channel (independent of 
the input) when pressed. The output test greatly aids in saving 
time during initial startup and/or troubleshooting. 
The output test level is factory set to approximately 75% of 
span but a custom test level may be ordered.

Reviews: