23
Energy Factors for BurstDR™ Stimulation Parameters
The following tables show energy factors according to various stimulation parameters for
BurstDR™ stimulation programs.
NOTE: Energy factors represent IPGs that provide 24 hours of daily stimulation using default
values for the Burst Frequency, Intra-burst Rate, and Pulse Width settings.
NOTE: In neurostimulation therapy, “dose” refers to the delivery of a quantity of energy to
tissue. A difference in “dose” in this context does not imply differences in expected
effectiveness response as it would with a drug. There is no demonstrated difference in safety
or effectiveness among these doses.
Table 10. Energy factors for various BurstDR stimulation parameters (350-ohm impedance)
Amplitude
(mA)
Intermittent Dosage (On Time/Off Time)
Continuous
Dosage
5 s/15 s
15 s/45 s 30 s/90 s 60 s/180 s 30 s/150 s 30 s/180 s 30 s/360 s
0.2
47
31
25
23
16
13
7
77
0.4
71
43
34
28
20
17
10
89
0.6
93
53
41
34
24
20
12
104
0.8
117
65
49
40
29
25
13
118
1.0
146
76
56
46
34
28
16
134
1.2
175
106
84
73
50
42
24
240
1.4
202
121
95
82
56
48
26
270
1.6
230
135
106
91
63
55
29
301
1.8
247
151
120
100
70
60
33
330
2.0
279
163
133
110
76
65
35
361
Table 11. Energy factors for various BurstDR stimulation parameters (500-ohm impedance)
Amplitude
(mA)
Intermittent Dosage (On Time/Off Time)
Continuous
Dosage
5 s/15 s
15 s/45 s 30 s/90 s 60 s/180 s 30 s/150 s 30 s/180 s 30 s/360 s
0.2
46
32
25
23
15
13
7
76
0.4
71
44
34
29
19
18
9
91
0.6
95
56
41
34
24
20
12
105
0.8
121
67
49
41
28
25
14
120
1.0
151
93
73
64
43
37
21
211
1.2
174
108
84
73
50
43
23
240
1.4
203
122
96
82
57
48
26
271
1.6
225
137
106
93
64
55
29
301
1.8
264
160
129
111
70
59
32
330
2.0
297
196
168
146
97
83
45
512