Instrument Clock Date/Time Synchronization
When the instrument is configured to transmit test results to a data manager or to receive patient,
operator, or cartridge lists from a data manager, the instrument will automatically synchronize the date
and time of its clock with the data manager’s time. This option eliminates the need to adjust the
instrument’s clock at the beginning and end of Daylight Saving Time.
To enable communication with data manager:
1. Install a customization profile with one of the following features selected:
-Test Records
-Operator List
-Patient List
-Cartridge List
2. Use NCi to configure the instrument for wired or wireless connectivity.
Note: When selecting only wireless connectivity ensure that wireless is enabled on
both the instrument and in the customization profile.
To ensure instrument time is synchronized with the data manager after a change in Daylight Saving Time,
power on the instrument and leave it on the Home screen until the correct time is displayed on the
Home screen before performing cartridge testing.
For instrument time synchronization to occur, the instrument must also have a wired or wireless
connection to the network hosting the data manager. The instrument may attempt to synchronize time
when it is communicating with a data manager to send results or receive lists. The instrument will also
attempt to synchronize time at power-on, so long as the Home screen is being displayed.
By default, instrument date/time synchronization is enabled, provided the conditions above are met, and
the option to Synchronize Clock with Data Manager is enabled. To disable the automatic
synchronization of date/time, run the Set Clock flow on the instrument and uncheck the box for
Synchronize Clock with Data Manager. Unchecking this box disables automatic synchronization of date/
time with the data manager and enables the option to manually adjust the date/time. The Synchronize
Clock with Data Manager check box is only available on the Set Clock screen when the instrument is
configured to communicate with a data manager.
24
i-STAT Alinity — System Operations Manual
Art: 745527-01 Rev. N
Rev. Date: 13-Oct-2022
Summary of Contents for i-STAT Alinity
Page 1: ......
Page 12: ...xii i STAT Alinity System Operations Manual Art 746982 01 Rev E Rev Date 02 Nov 2022 ...
Page 24: ...xxiv i STAT Alinity System Operations Manual Art 746983 01 Rev H Rev Date 11 Nov 2020 ...
Page 39: ......
Page 44: ...6 i STAT Alinity System Operations Manual Art 745526 01 Rev H Rev Date 04 Mar 2021 ...
Page 48: ...10 i STAT Alinity System Operations Manual Art 745527 01 Rev N Rev Date 13 Oct 2022 ...
Page 84: ...46 i STAT Alinity System Operations Manual Art 745531 01 Rev G Rev Date 04 Mar 2021 ...
Page 88: ...50 i STAT Alinity System Operations Manual Art 745529 01 Rev G Rev Date 04 Mar 2021 ...
Page 96: ...58 i STAT Alinity System Operations Manual Art 745530 01 Rev D Rev Date 16 May 2022 ...
Page 100: ...62 i STAT Alinity System Operations Manual Art 745578 01 Rev C Rev Date 04 Mar 2021 ...
Page 108: ...70 i STAT Alinity System Operations Manual Art 745525 01 Rev E Rev Date 04 Mar 2021 ...
Page 110: ...72 i STAT Alinity System Operations Manual Art 745532 01 Rev D Rev Date 04 Mar 2021 ...
Page 134: ...96 i STAT Alinity System Operations Manual Art 745524 01 Rev I Rev Date 02 Nov 2022 ...
Page 147: ......
Page 164: ...126 i STAT Alinity System Operations Manual Art 747079 01 Rev F Rev Date 28 Apr 2020 ...
Page 167: ......
Page 176: ...136 i STAT Alinity System Operations Manual Art 746298 01 Rev M Rev Date 13 Oct 2022 ...
Page 250: ...210 i STAT Alinity System Operations Manual Art 746300 01 Rev N Rev Date 02 Nov 2022 ...
Page 290: ...250 i STAT Alinity System Operations Manual Art 746300 01 Rev N Rev Date 02 Nov 2022 ...
Page 296: ...256 i STAT Alinity System Operations Manual Art 746304 01 Rev H Rev Date 29 Oct 2019 ...