Basic Functions
Options on this tab control access to training mode and provide a workspace to create training scenarios.
Training Mode
Allow Instrument to Enter
Training Mode
To deny access to the training path, unselect the check box. If access
to the training path is enabled, choose one of these options:
Require permissions
This is the default.
No permissions required
Select to allow access to the
training path regardless of
permissions.
Training Pathway Cartridges
Allow Expired Cartridges in
Training Pathway
Select this check box to allow expired cartridges to be used in the
training path.
Operator Direct Observation Checklist
Edit Observation Checklist
Select this option to create a checklist of items to display on the
i
‑
STAT Alinity instrument. Observation checklist is intended to be
used as a location on the instrument where a super user (trainer)
and an operator can see a comprehensive list of the skills needed to
remain competent. The super user (trainer) observes the operator
performing the skills and communicates to the POCC (administrator)
that the operator successfully completed the list.
A text box displays and up to 40 characters can be entered. To save,
click Add Observation. This will display another text box for an
additional item. When all items have been entered, specify an
identifier for the revision in the Checklist Revision text box.
Update Observation Checklist
Select this option after all the checklist items have been created and
a Checklist Revision value was supplied. The checklist items and
checklist revision information will display.
Training Scenarios
Add Training Scenario
Select this option to open a workspace to create training scenarios.
Be aware of the following:
•
For the patient information to be displayed on the i
‑
STAT Alinity
instrument, the PPID feature in Patient Settings must be
enabled.
•
For analyte values to display as critical, abnormal, or normal
values, reference and action ranges in Analyte Settings are
required.
206
i-STAT Alinity — System Operations Manual
Art: 746300-01 Rev. N
Rev. Date: 02-Nov-2022
Summary of Contents for i-STAT Alinity
Page 1: ......
Page 12: ...xii i STAT Alinity System Operations Manual Art 746982 01 Rev E Rev Date 02 Nov 2022 ...
Page 24: ...xxiv i STAT Alinity System Operations Manual Art 746983 01 Rev H Rev Date 11 Nov 2020 ...
Page 39: ......
Page 44: ...6 i STAT Alinity System Operations Manual Art 745526 01 Rev H Rev Date 04 Mar 2021 ...
Page 48: ...10 i STAT Alinity System Operations Manual Art 745527 01 Rev N Rev Date 13 Oct 2022 ...
Page 84: ...46 i STAT Alinity System Operations Manual Art 745531 01 Rev G Rev Date 04 Mar 2021 ...
Page 88: ...50 i STAT Alinity System Operations Manual Art 745529 01 Rev G Rev Date 04 Mar 2021 ...
Page 96: ...58 i STAT Alinity System Operations Manual Art 745530 01 Rev D Rev Date 16 May 2022 ...
Page 100: ...62 i STAT Alinity System Operations Manual Art 745578 01 Rev C Rev Date 04 Mar 2021 ...
Page 108: ...70 i STAT Alinity System Operations Manual Art 745525 01 Rev E Rev Date 04 Mar 2021 ...
Page 110: ...72 i STAT Alinity System Operations Manual Art 745532 01 Rev D Rev Date 04 Mar 2021 ...
Page 134: ...96 i STAT Alinity System Operations Manual Art 745524 01 Rev I Rev Date 02 Nov 2022 ...
Page 147: ......
Page 164: ...126 i STAT Alinity System Operations Manual Art 747079 01 Rev F Rev Date 28 Apr 2020 ...
Page 167: ......
Page 176: ...136 i STAT Alinity System Operations Manual Art 746298 01 Rev M Rev Date 13 Oct 2022 ...
Page 250: ...210 i STAT Alinity System Operations Manual Art 746300 01 Rev N Rev Date 02 Nov 2022 ...
Page 290: ...250 i STAT Alinity System Operations Manual Art 746300 01 Rev N Rev Date 02 Nov 2022 ...
Page 296: ...256 i STAT Alinity System Operations Manual Art 746304 01 Rev H Rev Date 29 Oct 2019 ...