Adjustments
The settings for result displays of base excess and eGFR (Estimated Glomerular Filtration Rate) are
determined in this tab.
BE Equation
Base excess of the extracellular fluid or standard base excess is the
concentration of titratable base minus the concentration of titratable
acid when titrating the average intracellular fluid (plasma plus interstitial
fluid) to an arterial plasma pH of 7.40 at PCO
2
of 40 mmHg at 37°C.
Excess concentration of base in the average ECF remains virtually
constant during acute changes in the PCO
2
and reflects only
nonrespiratory component of pH disturbances. Select either option:
Extra Cellular Fluid
Base Excess of Extracellular Fluid (BEecf)
BEecf = HCO
3
– 24.8 + 16.2 (pH – 7.4)
This is the default.
Blood
Base Excess of Blood (BEb)
BEb = (1 - 0.014*Hb) * [HCO
3
- 24.8 + (1.43 * Hb
+ 7.7) * (pH - 7.4)]
eGFR Equation
MDRD (default)
CKD-EPI 2009
eGFR Variants
Select one of these options:
Display both eGFR and Black/African American eGFR
Display only eGFR
Display only Black/African American eGFR
Note: Black/African American eGFR displays as eGFR-a on the
instrument screen.
When finished, choose from the options shown at the bottom of the screen:
•
Previous to go back to the previous screen
•
Discard to undo changes
•
Finish Later to create a draft
•
Next to open the next tab
188
i-STAT Alinity — System Operations Manual
Art: 746300-01 Rev. N
Rev. Date: 02-Nov-2022
Summary of Contents for i-STAT Alinity
Page 1: ......
Page 12: ...xii i STAT Alinity System Operations Manual Art 746982 01 Rev E Rev Date 02 Nov 2022 ...
Page 24: ...xxiv i STAT Alinity System Operations Manual Art 746983 01 Rev H Rev Date 11 Nov 2020 ...
Page 39: ......
Page 44: ...6 i STAT Alinity System Operations Manual Art 745526 01 Rev H Rev Date 04 Mar 2021 ...
Page 48: ...10 i STAT Alinity System Operations Manual Art 745527 01 Rev N Rev Date 13 Oct 2022 ...
Page 84: ...46 i STAT Alinity System Operations Manual Art 745531 01 Rev G Rev Date 04 Mar 2021 ...
Page 88: ...50 i STAT Alinity System Operations Manual Art 745529 01 Rev G Rev Date 04 Mar 2021 ...
Page 96: ...58 i STAT Alinity System Operations Manual Art 745530 01 Rev D Rev Date 16 May 2022 ...
Page 100: ...62 i STAT Alinity System Operations Manual Art 745578 01 Rev C Rev Date 04 Mar 2021 ...
Page 108: ...70 i STAT Alinity System Operations Manual Art 745525 01 Rev E Rev Date 04 Mar 2021 ...
Page 110: ...72 i STAT Alinity System Operations Manual Art 745532 01 Rev D Rev Date 04 Mar 2021 ...
Page 134: ...96 i STAT Alinity System Operations Manual Art 745524 01 Rev I Rev Date 02 Nov 2022 ...
Page 147: ......
Page 164: ...126 i STAT Alinity System Operations Manual Art 747079 01 Rev F Rev Date 28 Apr 2020 ...
Page 167: ......
Page 176: ...136 i STAT Alinity System Operations Manual Art 746298 01 Rev M Rev Date 13 Oct 2022 ...
Page 250: ...210 i STAT Alinity System Operations Manual Art 746300 01 Rev N Rev Date 02 Nov 2022 ...
Page 290: ...250 i STAT Alinity System Operations Manual Art 746300 01 Rev N Rev Date 02 Nov 2022 ...
Page 296: ...256 i STAT Alinity System Operations Manual Art 746304 01 Rev H Rev Date 29 Oct 2019 ...