1.12 - i-STAT Cartridges
Overview
i
‑
STAT cartridges contain test reagents which are located on sensors on the top of the cartridge. The
instrument and cartridge work together to generate a clinically meaningful result.
The cartridges are designed for use with the i
‑
STAT Alinity instrument.
Note:
•
Not all cartridges are available in all regions. Check with your local representative for
availability in specific markets. Scanning a cartridge that is unavailable will result in a
pop-up window that displays Invalid Cartridge Type.
•
Customization may affect analyte availability.
•
For access to cartridge-specific Instructions for Use (IFU) and analyte-specific CTI
Sheets, access the Support page on the Abbott Point of Care website
.
Cartridge Specifications
Shelf Life: Refrigerated at 2ºC to 8ºC (35ºF to 46ºF) until expiration date. Refer to the cartridge box for
room temperature storage requirements.
Preparation for Use: Individual cartridges may be used after standing five minutes at room temperature.
An entire box of cartridges should stand at room temperature for one hour. All cartridges should be used
immediately after opening pouch. If the pouch has been punctured, the cartridge should not be used.
Cartridges stored in refrigerator:
•
Temperature must be 2-8°C (35-46°F)
•
Cartridges expire on the date printed on the pouch
Cartridges stored at room temperature:
•
Temperature must be 18-30°C (64-86°F)
•
Once cartridge is at room temperature the expiration date changes
•
Cartridge should not be returned to the refrigerator once out for more than 5 minutes.
85
i-STAT Alinity — System Operations Manual
Art: 745524-01 Rev. I
Rev. Date: 02-Nov-2022
Summary of Contents for i-STAT Alinity
Page 1: ......
Page 12: ...xii i STAT Alinity System Operations Manual Art 746982 01 Rev E Rev Date 02 Nov 2022 ...
Page 24: ...xxiv i STAT Alinity System Operations Manual Art 746983 01 Rev H Rev Date 11 Nov 2020 ...
Page 39: ......
Page 44: ...6 i STAT Alinity System Operations Manual Art 745526 01 Rev H Rev Date 04 Mar 2021 ...
Page 48: ...10 i STAT Alinity System Operations Manual Art 745527 01 Rev N Rev Date 13 Oct 2022 ...
Page 84: ...46 i STAT Alinity System Operations Manual Art 745531 01 Rev G Rev Date 04 Mar 2021 ...
Page 88: ...50 i STAT Alinity System Operations Manual Art 745529 01 Rev G Rev Date 04 Mar 2021 ...
Page 96: ...58 i STAT Alinity System Operations Manual Art 745530 01 Rev D Rev Date 16 May 2022 ...
Page 100: ...62 i STAT Alinity System Operations Manual Art 745578 01 Rev C Rev Date 04 Mar 2021 ...
Page 108: ...70 i STAT Alinity System Operations Manual Art 745525 01 Rev E Rev Date 04 Mar 2021 ...
Page 110: ...72 i STAT Alinity System Operations Manual Art 745532 01 Rev D Rev Date 04 Mar 2021 ...
Page 134: ...96 i STAT Alinity System Operations Manual Art 745524 01 Rev I Rev Date 02 Nov 2022 ...
Page 147: ......
Page 164: ...126 i STAT Alinity System Operations Manual Art 747079 01 Rev F Rev Date 28 Apr 2020 ...
Page 167: ......
Page 176: ...136 i STAT Alinity System Operations Manual Art 746298 01 Rev M Rev Date 13 Oct 2022 ...
Page 250: ...210 i STAT Alinity System Operations Manual Art 746300 01 Rev N Rev Date 02 Nov 2022 ...
Page 290: ...250 i STAT Alinity System Operations Manual Art 746300 01 Rev N Rev Date 02 Nov 2022 ...
Page 296: ...256 i STAT Alinity System Operations Manual Art 746304 01 Rev H Rev Date 29 Oct 2019 ...