Setup Guide: CONNECT Universal Gateway With ID NOW
TM
(Direct-Connect)
Abbott Rapid Diagnostics Informatics, Inc. (ARDx Informatics)
D10036883 REV AA
Page 7 of 26
February 1, 2021
© 2021 Abbott. All rights reserved.
All trademarks referenced are trademarks of either the Abbott group of companies, or their respective owners.
Any photographs displayed are for illustrative purposes only.
Configuration
In order for the
first
ID NOW
TM
instrument to communicate with the CONNECT Universal Gateway, additional
configuration steps are required, as follows:
Step
Action
Illustration
1
VERIFY
that the first ID NOW
TM
device is powered on;
TURN
ON
the CONNECT Universal Gateway.
Note: The CONNECT Universal Gateway will
automatically switch on once attached to power. If
this does not occur, press the Power button.
2
CHECK
the status lights on the CONNECT
Universal Gateway
Observe and check the three device status
lights
1, 2
and
3.
If a solid
Status 1
light does not appear within
two minutes, restart the Universal Gateway by
pressing and holding down the Power button
for five seconds.
The
Status 2
light will flash while the device
attempts to connect.
The
Status 3
light will remain solid green
when a successful connection is established
with RALS LiNK
TM
or Sympheos
TM
.
3
POWER ON
the first ID NOW
TM
device.
Ssubsequent steps will be completed on the first ID
NOW
TM
device.