7.16 Flow range
Pressure Display
Speed Step Resolution
Flow Range
Flow Limit Sensitivity
Flow Recorder Speed*
Set RPM
Exit
Menu
Active
Inactive
Set RPM
Set RPM
Set RPM
Set RPM
Set RPM
Set RPM
Menu
Menu
Menu
Menu
Menu
Menu
Normal
(100 rpm)
Fine
(50 rpm)
Set RPM
Menu
R
Normal
Sensitive
Set RPM
Menu
Next
Normal
(0 – 10 lpm)
Reduced
(0 – 3 lpm)
Slow
(2 min)
Fast
(20 sec)
Set RPM
Menu
L
Select
Enter
Return
* Monitor only
Extended Menu
Set RPM
Menu
[ |Confirm]
[Execute|Cancel]
Set RPM
Menu
[ |Confirm]
[Start|Cancel]
Copy Data*
Management*
Support Type
Language
Data Logger*
Figure 44: Monitor MENU Structure
Flow Range
The System may be used to treat a variety of conditions to accommodate different ranges
of flow. Under different conditions, the System may be operated in one of two flow ranges
0 – 10 LPM and 0 – 3 LPM. Changing the flow range will adjust the range of the flow
display bar, as well as the default limit for the maximum flow alarm if the high limit is more
than 3.0 LPM. The most common use of the low flow range is for display of flow during
use of the smaller flow transducer designed to be placed on ¼” ID tubing.
To change between the two ranges, depress the
MENU
button until the
EXTENDED
MENU
option is displayed, then use the
MENU
key to scroll through the menu options until
the
FLOW RANGE
option is displayed. Select between the two options using the arrow
keys.