CentriMag
TM
Circulatory Support System Operation Manual (US)
Page 69 of 126
© 2019 Abbott – Document No PL-0047, Rev 11 (December 2019)
To change the
MAXIMUM
or
MINIMUM PRESSURE ALERT
settings for the P1 pressure
transducer, depress the
MENU
keypad to display PRESSURE MENU followed by ENTER.
Select either the
MAX
or
MIN P1 ALERT
options. Use the
DECREASE
or
INCREASE
arrow keys to decrease or increase the alert setting.
7.13 Setting the Max or Min Pressure Alert Settings for the P2 Transducer
Pressure Menu
Minimum P1 Alert
Maximum P1 Alert
Minimum P2 Alert
Maximum P2 Alert
Pressure Calibration
Set RPM
Exit
Menu
Increase
Decrease
Set RPM
Set RPM
Set RPM
Set RPM
Set RPM
Set RPM
Menu
Menu
Menu
Menu
Menu
Menu
Increase
Decrease
Increase
Decrease
Increase
Decrease
Return
[New|Cal P1]
[Previous|Cal P2]
Figure 40: Monitor MENU Structure
Setting the Max or Min Pressure Alert Settings for the P2 Transducer
To change the
MAXIMUM
or
MINIMUM PRESSURE ALERT
settings for the P2 pressure
transducer, depress the
MENU
keypad to display PRESSURE MENU followed by ENTER.
Select either the
MAX
or
MIN P2 ALERT
options. Use the
DECREASE
or
INCREASE
arrow keys to decrease or increase the alert setting.
7.14 Activating the Pressure Monitoring System (Pressure Display)
1. Install the Pressure Probes into the Pressure Probe Cables and connect the Cables
into the mating connectors P1 and P2 on the back of the Console. For operation of a
single pressure probe, connect the cable into the P1 connector on the back of the
Console.
WARNING
If there is a failure to obtain pressure data confirm that the pressure transducer
and cable connection are fully seated. Additional troubleshooting may include
recalibrating, or disconnecting and reconnecting the connections.