CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Reticulocyte Package
Quality Control Guide
Section 14
Quality Control Specimens
Always mix and handle commercial control material according to the directions
given in the package insert. Pay particular attention to the following:
• Store the CELL-DYN Retic-RITE™ controls in the refrigerator at 2°– 8° C.
Store in a suitable location in the refrigerator, away from the door if it is
opened frequently.
• Carefully warm the CELL-DYN Retic-RITE™ controls prior to mixing,
according to the directions given in the package insert. Proper mixing is
essential for accurate results.
• Mix the CELL-DYN Retic-RITE™ control vials gently by hand to
thoroughly resuspend the control material. Do not use automatic mixers to
resuspend the control material.
• Check the open-vial stability dating given on the package insert and do not
use the products longer than is recommended or results may be
compromised.