14-42
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Reticulocyte Package
Retic Menu Options
Section 14
The data displayed on the screen are the kinetic data for the Reticulocyte specimen
from the last run, displayed in a tabular format. (See Figure 14.23.) The total count,
data acquisition intervals, and rate per second are displayed.
NOTE:
Time is displayed on the X-axis. Raw count is displayed on the Y-axis.
When the
[RETIC CNT GRAPH]
key is pressed, the rate per second data are displayed
as a graph. (See the following figure.) The kinetic data and graph information are
useful when troubleshooting problems with the reticulocyte parameter. A printout
of the Reticulocyte Count Rate may be obtained by pressing the
[PRINT]
key when
the desired format is displayed on the screen.
Figure 14.24 Reticulocyte Count Rate Screen (Graphic Format)
Clear Faults
When the
[CLEAR FAULTS]
key is pressed, the Analyzer returns to the READY state
if the corrective action that was taken has resolved the problem. If the corrective
action did not correct the problem, the fault status does not change.
NOTE:
Only operator-correctable faults can be cleared with the
[CLEAR FAULTS]
key.
RETIC
CNT DATA
DIAGNOSTICS MENU
Ready
Aug 02 2001
Operator ID
Retic Seq#
14:32
R0482
TS
RETURN
4405.9
0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5
RETIC CNT GRAPH
3855.2
3304.5
2753.7
2203.0
1652.2
1101.5
550.7