![Abbott CELL-DYN 3200 System Operator'S Manual Download Page 397](http://html1.mh-extra.com/html/abbott/cell-dyn-3200/cell-dyn-3200_system-operators-manual_450985397.webp)
CELL-DYN
®
3200 System Operator’s Manual
9-27
9140181H—October 2001
Section 9
Service and Maintenance
Semiannual Maintenance Procedures
Printer Cleaning
Every six months (or after about 300 hours of operation) turn the printer OFF,
disconnect the power cord and use a clean, dry cloth to dust the area around the
carriage shaft and platen or ink cartridge (depending on type of printer). Be sure to
remove any loose particles of paper. Do not use solvents or strong detergents on the
cabinet.
Refer to the printer manual that accompanied your printer for detailed cleaning and
maintenance instructions.