6-54
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Calibration Procedures
Manual Calibration Worksheets
Section 6
Worksheet 4
Mode to Mode Calibration Difference
Instrument:
Date:
Operator:
1. In column 1, enter the % Diff from column 5 of Worksheet 4 (disregard the sign).
2. If the % Diff exceeds the limit in column 4, DO NOT CALIBRATE. Call the Abbott Hematology
Customer Support Center for assistance.
Mode to Mode Calibration Difference
(Calculate All Factors To Three Decimal Places)
(1)
Closed
Mode Mean
–
(2)
Open Mode
Mean
/
(3)
Open Mode
Mean
x 100 =
(4)
% Diff
WOC
–
/
x 100 =
NOC
–
/
x 100 =
RBC
–
/
x 100 =
HGB
–
/
x 100 =
MCV
–
/
x 100 =
PLT
–
/
x 100 =
Worksheet 5
Mode to Mode Calibration Criteria
(1)
% Diff
(2)
Lower Limit
Cal Not Required
(3)
Calibration Range
Cal Required
(4)
Upper Limit
Do Not Cal
(5)
Cal? Y/N
WOC
<2.0%
>2.0% but <10%
>10%
NOC
<2.0%
>2.0% but <10%
>10%
RBC
<2.0%
>2.0% but <10%
>10%
HGB
<2.0%
>2.0% but <10%
>10%
MCV
<2.0%
>2.0% but <10%
>10%
PLT
<3.0%
>3.0% but <15%
>15%