CELL-DYN
®
3200 System Operator’s Manual
5-101
9140181H—October 2001
Section 5
Operating Instructions
Prolonged Shutdown
If the shutdown time is expected to exceed two weeks, follow the procedure below:
1. In the
MAIN MENU
, press
[SPECIAL PROTOCOLS]
2. Press
[AUTO CLEAN]
to clean the instrument.
3. Press
[PREPARE FOR SHIPPING]
. Follow the instructions on the screen.
4. When the procedure is completed, turn the power OFF.
5. Remove the following tubing:
• Tubing in the Normally Closed Valves on the Flow Panel (refer to Figure
1.5).
• Tubing under the Peristaltic Pump.
6. Record this maintenance in your maintenance log.
CAUTION:
Do not forget to reinsert the tubing securely in the Normally
Closed Valves and under the peristaltic pump before turning the instrument
back ON. Refer to
Installation Procedures and Special
.
For the SL model, also refer to the Prolonged Shutdown Procedure in
.
If the instrument will be shipped, refer to the instructions in