5-100
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Operating Instructions
Daily Shutdown Procedure
Section 5
Peristaltic Pump Tubing
IMPORTANT:
By removing the tubing under the Sample Transfer Peristaltic
Pump, the life of the tubing is significantly prolonged. The tubing should be
removed if the instrument will be idle for more than 24 hours and reinserted prior
to operating the instrument.
Short Term Shutdown
If the shutdown time is expected to be 72 hours or less, follow the procedure below:
1. In the
MAIN MENU
, press
[SPECIAL PROTOCOLS]
.
2. If the instrument is to be cleaned first, press
[AUTO CLEAN]
, otherwise go to
step 3.
3. Press
[DAILY SHUTDOWN]
.
4. Leave power ON.
5. Tubing should be removed from under the Peristaltic Pump.
Intermediate-term Shutdown
If the shutdown time is expected to exceed 72 hours (but less than 2 weeks), follow
the procedure below:
1. With the system still ON, perform any maintenance that is due.
2. In the
MAIN MENU
, press
[SPECIAL PROTOCOLS]
.
3. Press
[AUTO CLEAN]
to clean the instrument.
4. Press
[DAILY SHUTDOWN]
.
5. When the
'$,/<6+87'2:1
cycle is finished, turn power OFF in the
following order:
a. System (includes Analyzer, Data Module, and Sample Loader)
b. Display Monitor
c. Printer
NOTE:
In an emergency situation, turn OFF power in any order as quickly
as possible.
6. Remove the tubing from under the Sample Transfer Peristaltic Pump.