5-98
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Operating Instructions
Alerts and Indicators
Section 5
3 Consecutive Flow Errors
If the Sample Loader is being used and the instrument detects one of the following:
1. Three consecutive RBC flow errors.
2. Three consecutive WOC flow errors.
3. Three consecutive NOC flow errors.
The Sample Loader halts at the end of the cycle and the Bulletin line displays the
appropriate message:
1.
3 CONSECUTIVE RBC FLOW ERRORS.
2.
3 CONSECUTIVE WOC FLOW ERRORS.
3.
3 CONSECUTIVE NOC FLOW ERRORS.
Sampling Errors
The message
SAMPLING ERROR-INCOMPLETE ASPIRATION
is displayed on
the Bulletin line if insufficient sample was detected during aspiration.
SAMPLING
ERR
is displayed on the screen and
SAMPLING ERR
is printed on the graphics
report to the right of the MCHC. The same message is printed to the right of the
WBC on the pre-printed ticket and printed above the list of parameters on the blank
ticket.
3 Consecutive Sampling Errors
If the Sample Loader is being used and the Instrument detects three consecutive
incomplete aspirations, the Sample Loader halts at the end of the cycle and the
Bulletin line displays the message
3 CONSECUTIVE INCOMPLETE ASPIRATIONS.