CELL-DYN
®
3200 System Operator’s Manual
5-25
9140181H—October 2001
Section 5
Operating Instructions
Figure 5.9 QC Range Entry Screen
Range Entry
1. Select a file from the
QC SET UP
screen by using the Arrow keys on the
keyboard to move the cursor into the desired file.
2. Press
[QC LIMITS]
to display the
QC RANGE ENTRY
screen for the selected file.
3. Use the Arrow keys on the keyboard to move the cursor to the desired entry
field.
4. Type the numbers and press the Enter key on the keyboard to save each entry
and advance the cursor to the next entry field.
5. Repeat step 4 until all entries have been made.
6. If desired, press
[PRINT]
to obtain a printout of the entered values.
7. Press
[RETURN]
to save the entries and return to the
QC SET UP
screen.
NOTE:
When the entries are saved by pressing the
[RETURN]
soft key, the
software checks to see if any entries would result in the upper limit
being less than the lower limit. If this situation occurs, the limits are
automatically reversed. The bulletin line displays the message:
LIMITS WERE EXCHANGED TO MAKE UPPER > LOWER