5-20
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Operating Instructions
Set Up Instructions
Section 5
To Delete Entries
When the log is full (the log holds 10 entries), the oldest entry must be deleted to
create space for a new entry. Abbott suggests that, for documentation purposes, the
log be printed when it is full.
1. Press
[REAGENT LOG]
to display a
REAGENT LOG
screen. If necessary, press
a soft key to display the appropriate log.
2. Move the cursor to the oldest entry in the log.
3. Press
[DELETE ENTRY]
. The
[COMPLETE DELETION]
and
[RESTORE ENTRY]
keys are displayed.
4. Press
[COMPLETE DELETION]
to delete the selected entry.
5. If desired, a new entry may then be made as described above.
NOTE:
New entries may also be made by typing over old entries without
deleting them.
6. Press
[RETURN]
to return to the main
SET UP MENU
screen.
Figure 5.6 QC Set Up Menu Screen