![Abbott CELL-DYN 3200 System Operator'S Manual Download Page 130](http://html1.mh-extra.com/html/abbott/cell-dyn-3200/cell-dyn-3200_system-operators-manual_450985130.webp)
3-36
CELL-DYN
®
3200 System Operator’s Manual
9140181K—July 2002
Principles of Operation
Operational Messages and Data Flagging
Section 3
Interpretive Messages
Interpretive messages appear only on the graphics report and are generated when the
numeric limits entered in the Patient Limit Sets are exceeded. (see
for an explanation).
These messages are printed only when the Interpretive Report option is selected on
the
CUSTOMIZE
REPORT
SCREEN
. The Interpretive messages are summarized below.
WBC Messages
RBC Messages
PLT Messages
Message
Cause
Leukopenia
result exceeds the lower limit for WBC
Leukocytosis
result exceeds the upper limit for WBC
Neutropenia
result exceeds the lower limit for Neutrophil absolute
number
Neutrophilia
result exceeds the upper limit for Neutrophil absolute
number
Lymphopenia
result exceeds the lower limit for Lymphocyte
absolute number
Lymphocytosis
result exceeds the upper limit for Lymphocyte
absolute number
Monocytosis
result exceeds the upper limit for Monocyte absolute
number
Eosinophilia
result exceeds the upper limit for Eosinophil absolute
number
Basophilia
result exceeds the upper limit for Basophil absolute number
Message
Cause
Anemia
result exceeds the lower limit for RBCs
Polycythemia
result exceeds the upper limit for RBCs
Microcytic RBC
result exceeds the lower limit for MCV
Macrocytic RBC
result exceeds the upper limit for MCV
Hypochromic
result exceeds the lower limit for MCHC
Hyperchromic
result exceeds the upper limit for MCHC
Anisocytosis
result exceeds the upper limit for RDW
Message
Cause
Thrombocytopenia
result exceeds the lower limit for PLTs
Thrombocytosis
result exceeds the upper limit for PLTs
Microcytic PLT
result exceeds the lower limit for MPV
Macrocytic PLT
result exceeds the upper limit for MPV