3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURESVP - 13.1 Lead Bubble Check/Adjustment
3 - 154
3
VP - 13.1 Lead Bubble Check/Adjustment
Purpose: Perform this procedure to verify and adjust Lead Bubble on E Loaders with old style (clear
plastic) Wash Blocks.
1. Verify proper connection of sample line to Sample Needle.
a. Slide the sleeve off of the sample line/Sample Needle connection.
b. Verify the sample line is properly connected to the Sample Needle.
c. Reseat the sleeve.
2. Verify sample tubing from Blood In-line Sensor to Solenoid 96 is not crimped.
3. Put Sample Loader in Manual Mode.
a. Press and Hold MAN and START.
b. Turn Power ON, hold keys down until BEEPING stops.
4. Clean the sample path:
a. Place tube with enzymatic cleaner under Sample Needle.
b. Lower Sample Needle into tube of cleaner.
c. Aspirate 1/2 tube cleaner by manually opening Solenoid 96 and 92.
Cleaner drawn through Sample Needle, Shear Valve and deposited in Waste Chamber 2.
Special Tools
DESCRIPTION
QTY
PN
None