CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
TROUBLESHOOTING IP - 1 Blood Detected In Sample Line, 3 of 3
1 - 12
1
IP - 1 Blood Detected In Sample Line, 3 of 3
CPU Main Bd.
Blood In-Line Sensor =
+5.0V With Blood <0.2V
Without (See Figure)
?
Replace Blood In-Line Sensor Bd. (
See Interrelationship Matrix -
See Interrelationship Matrix -
Y e s
2
N o
[NON-CE MARK]
TP
SL-124D
[CE MARK]
CN
5
Pin 23
Blood
In-Line
Pin 17 Blood In-Line