28 Oct
1997 21:57
CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURESVP - 3.10 X-Axis Solenoids to Platen Channel Check
3 - 43
3
VP - 3.10 X-Axis Solenoids to Platen Channel Check
Purpose: Perform this procedure to verify Platen Channels are parallel to X-Axis Guide Rods.
1. Put Sample Loader in Manual Mode.
a. Press and Hold MAN and START.
b. Turn Power ON, hold keys down until BEEPING stops.
2. Press (
←
X
→)
keys, move the solenoids along their full length of travel.
3. Visually confirm the following:
• The solenoids complete the movement without touching the Platen.
• The
gap between each solenoid and the Platen Channel is approximately the same all along
the length of travel.
4. IF Check is OK, go to
“VP - 3.12 Platen to Laser Shield Check” (Page 3-47)
.
5. IF Check Fails, go to
“VP - 3.11 X-Axis Solenoids to Platen Channel Adjustment” (Page 3-45)
Special Tools
DESCRIPTION
QTY
PN
None