10-64
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Troubleshooting
Chapter 10
Table 10.2
General Fault Conditions (Continued)
Auto-Sampler alarm condition <u>; see DIAG/BL
– – – – – Probable Cause – – – –
– – – Corrective Action – – – – –
1.
Sample Loader rack
1.
Clear the rack movement
movement is impeded.
obstruction and/or clean
the Sample Loader racks
and trays.
Press the [CLEAR FAULT]
soft key on the
to reinitialize the Sample
Loader and bring the Data
Station to the READY state.
OR
REPORT] soft key. Then
press the [CLEAR FAULTS]
soft key to reinitialize the
Sample Loader and bring the
Data Station to the READY
state. (It is not necessary to
press the Sample Loader
RESET key.)
2.
The rack and/or position
2.
Check and/or replace the
ID label is unreadable.
rack ID/position ID label.
3.
The specimen bar code
3.
Ensure that your bar code
label is the same as an
labels are not replicates of
ID label used on the rack
the rack and position ID
or slots.
labels.