10-18
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Troubleshooting
Chapter 10
When the [VOLTAGE READINGS] key is pressed, the voltage and
vacuum/pressure values from a test point, measured at the moment when
the key was pressed, is displayed. (See Figure 10.15.) The following
additional soft key labels are displayed:
FINISH SELECT
SELECT
These two keys are for service use only.
The data provided by the VOLTAGE READINGS Screen can be useful
in determining if a problem is caused by a hardware malfunction.
Figure 10.15:
Voltage Readings Screen
VOLTAGE
READINGS
FINISH
SELECT
SELECT
DIGITAL
READINGS
VOLTAGE
READINGS
MORE
MAIN
DIAGNOSTICS MENU
Ready
Jan 28 1993
Operator ID
Sequence #
12:40
734
3143
Arrow keys to move around, SELECT key to select, FINISH SELECT key to go
DCM:
slf-tst f/ DAC
:0.00
99mv refrence
:0.08
15v/2 pwr sp
:7.49
5v pwr sp
:5.14
slf-test ramp
:9.99
9.901v refrnce
:9.86
-15v/2 pwr sp
:-7.55
SPM:
WBC threshold
:0.86
SPM tst f/ DAC
:0.00
WBC ch 1 peak
:0.00
RBC peak
:0.00
RBC threshold
:0.57
5v supply
:5.12
WBC ch 2 peak
:0.00
RBC intgral
:0.00
PLT L thrshld
:0.53
10v refrence
:9.97
WBC ch 3 peak
:0.00
PLT peak
:0.04
PLT H thrshld
:9.97
-10v refrence
:-10.0
WBC ch 4 peak
:0.00
MAM:
ch 1 offset
:0.20
ch 5 offset
:-2.06
electrode v/2
:0.41
laser ref
:0.00
ch 2 offset
:0.97
ch 6 offset
:-3.00
apt crrnt set
:-0.05
ch 3 offset
:0.64
ch 3-Vdyn/100
:6.14
tst puls H set
:-0.05
ch 4 offset
:0.27
ch 4-Vdyn/100
:5.65
tst puls L set
:-0.00
VPM:
press 1 psi
:12.0
press 3 psi
:4.88
vac 1 in. Hg
:11.8
pos ref pres
:5.00
press 2 psi
:9.01
vac 2 in. Hg
:11.7
neg ref pres
:-5.07
FCM:
HGB output
:5.53