10-10
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Troubleshooting
Chapter 10
Figure 10.7:
WBC Count Rate Graph
When the [CLEAR FAULTS] key is pressed, the Analyzer returns to the
READY state if the corrective action taken resolved the problem. If the
corrective action did not correct the problem, the fault status does not
change.
NOTE: Only operator-correctable faults can be cleared with the
[CLEAR FAULTS] key.
When the [RAW DATA SUMMARY] key is pressed, detailed
information pertaining to the last cycle run is displayed. An example of
the
useful information for the operator, the
The information on metering times may be used to assist in
troubleshooting chronic clog or flow error messages. The HGB reference
and sample readings may be used to assist in troubleshooting erratic or
imprecise HGB results.