Precautions, Limitations and Hazards
Chapter 8
Sample Loader Precautions
•
Samples that will be run on the Sample Loader should be collected
in 13 x 75 mm tubes. The recommended tube is a 13 x 75 mm tube
with a hemoguard closure that draws 3 mL of blood.
•
The safety cover has an interlock switch that prevents Sample
Loader operation when the cover is not in place. Lifting the cover
while the Sample Loader is operating causes an immediate
Emergency Stop condition. Do not bypass the interlock switch to
operate the Sample Loader without the safety cover.
•
Do not allow the tube racks to soak in water. Do not use an
automated washing system that operates at elevated temperatures
as this may damage the racks.
•
Clean the Sample Loader needles prior to touching them by
sampling with a tube containing enzymatic cleaner followed by an
Auto-Clean procedure. Wear gloves, lab coat and suitable eye
protection and follow other biosafety practices as specified in the
OSHA Bloodborne Pathogen Rule or other equivalent biosafety
procedures.
•
Liquid spills in the rack drive mechanism are a potential reason for
failure of the rack to advance. Liquid spills that flow in the Sample
Loader Control Panel could cause operational failure. If a spill
occurs, turn the Sample Loader off immediately and notify the
Customer Support Center for further assistance at 1 (800) CELL
DYN.
WARNING: Potential Biohazard. The needle is sharp and
potentially contaminated with infectious materials. Handle with
care.