8-2
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Precautions, Limitations and Hazards Chapter 8
•
The following space should be available to ensure proper
ventilation:
•
Behind the Analyzer and Data Station: 6" (15 cm)
•
Left of the Analyzer: 6" (15 cm)
•
Above the Analyzer and Data Station: 6" (15 cm)
•
Right of the Data Station: 6" (15 cm)
(As long as the Data Station is located to the right of the Analyzer,
no space is required between the two units.)
•
Care should be taken to prevent blocking of the air vents on the left
side of the Analyzer, on the back of the Analyzer, and on the back
of the Data Station.
•
Pay special attention to the path of the printer paper in the event
that the printer is placed on top of the Data Station. (Do not place
the printer on top of the Analyzer.) Refer to
more details.
Electrical Safety Precautions
•
Do not disconnect any electrical connection while the power is ON.
•
For continued protection from electric shock, use only approved
power cords such as those supplied and connect power cords only
to properly grounded outlets.
WARNING: Electrical Shock Hazard. Do not remove any
instrument panel cover that is securely fastened in place by
screws. The potential for electrical shock is increased whenever
these panels are removed.
•
Replace only the externally accessible, labeled
the power cord connector. Use a replacement fuse only of the
specified type and electrical rating. Refer to