CELL-DYN
3000 System Operator’s Manual
5-95
9140240E — May 1995
Chapter 5 Operating Instructions
6.
After the information for the sample is entered, place the sample in
a Sample Loader rack. Subsequent samples should be placed in the
racks in the order in which the sample information is entered into
the Work List.
NOTE: The last group of samples should be placed in an end
rack so the Sample Loader stops automatically when all the
samples have been processed.
7.
When all entries have been made, place the racks in order in the
Sample Loader and install the safety cover.
8.
Press [RETURN] to return to the
9.
Press [SPECIMEN TYPE] followed by [PATIENT].
10.
If necessary, press [CHANGE SAMPLER] to select the Closed
mode.
11.
Press the START key on the Sample Loader.
NOTE: Samples may be run with the
displayed. As the samples are processed, the status of each
sample is displayed in the <STATUS> field on the Work List.
Additional entries may be made to the Work List while
processing takes place.
12.
The samples are automatically processed in the order they are
placed in the racks. The Sample Loader automatically stops when
processing is finished if the last samples were placed in an end
rack.
Running STAT Samples
It is recommended that STAT samples be run in the Open Mode as
directed in the following procedure:
1.
Press the PAUSE key on the Sample Loader. The Sample Loader
stops after processing of the sample in progress is completed.
2.
, press [RETURN] to return to the
3.
, press [CHANGE
SAMPLER] to select the Open mode.
NOTE: The Work List is not active in the Open mode on the
CELL-DYN 3000SL.
4.
Type the Specimen ID in the <NEXT ID> field and press the Enter
key on the keyboard to save it.