5-66
CELL-DYN
3000 System Operator’s Manual
9140240E — May 1995
Operating Instructions Chapter 5
Procedure Resistant RBC
1.
, press [CHANGE SAMPLER]
to select the Open Mode.
2.
Press [SPECIMEN TYPE] followed by [RESISTANT RBC] to
select the Resistant RBC cycle.
3.
Enter the appropriate Patient ID information on the line labeled Res
RBC.
4.
Run the sample in the Open mode.
5.
When the cycle is complete, press [SPECIMEN TYPE] followed
by [PATIENT] to return to the normal RUN cycle.
The [CHANGE SAMPLER] key is used to select the Open or Closed
mode of operation. When the key is pressed, the mode changes from the
mode currently selected to the other operating mode. The Status Box
displays the message:
SELECTING OPEN MODE or SELECTING CLOSED MODE
When the cursor is positioned on the word AUTO (at the end of the
<NEXT ID> entry field), the key label changes to [TOGGLE AUTO
ID]. This key enables (word AUTO is highlighted) or disables the auto
increment feature. Refer to the
The [PRINT TICKET] key is used to print a report on a ticket when the
Ticket Printer is connected to the Data Station. The report is printed on
the type of ticket that is selected from the
The [PRINT REPORT] key is used to print a graphics report when the
Graphics Printer is connected to the Data Station.
The [MAIN] key is used to return to the
CUSTOMIZE
REPORT
CHANGE
SAMPLER
TICKET
REPORT
MAIN