CELL-DYN
3000 System Operator’s Manual
5-27
9140240E — May 1995
Chapter 5 Operating Instructions
Procedure Customize QC Log Display
(Standard Groups)
1.
Select a file from the
cursor to the desired file.
2.
Press [CUSTOMIZE DISPLAY] to display the
3.
Press [STANDARD GROUPS] to display the
and key labels for Standard Groups.
4.
Use the Arrow keys on the keyboard to move the cursor to the
desired group (1–4) location.
NOTE: This number indicates the order in which the group of
parameters will be displayed (Group 1 on the first screen, Group
2 on the second, etc.).
5.
Press the soft key corresponding to the desired parameter group.
This group is displayed in the position indicated by the cursor.
6.
Repeat steps 4 and 5 until all desired groups have been selected.
7.
If desired, press the Print Screen key on the keyboard to obtain a
printout of the configuration.
8.
Press [RETURN] to return to the
9.
Repeat this procedure to select standard groups for other QC logs.
The [LATEX SET] key is available from the
for Standard Groups. This key is used to customize the
QC file to store information generated by latex particles. Information is
stored for each of the four angles of scatter which are used to determine
the differential. This key is intended for Abbott Service personnel.