Operating Instructions Chapter 5
to the following numbered options:
1.
LOWER/UPPER LIMITS
The Lower and Upper Limits determine which patient results will
be used in the X-B Moving Average calculation. Results that fall
outside these limits are automatically excluded from the X-B
calculations. These limits should be set widely to exclude grossly
abnormal samples that would bias the calculation but should
include at least 95% of the patient results.
2.
TARGET VALUE
The Target Value for X-B Analysis is similar to the assay value for
a commercial control. It is derived from the patient population that
is analyzed on the instrument.
NOTE: The default (preassigned) Target Values on the
Target Values. Dr. Bull's Target Values are included in the
3.
ACTION LIMIT
The Action Limit is the acceptable limit of variation around the
target value.
Procedure X-B Setup
1.
, press [X-B SETUP] to display the
2.
Use the Arrow keys on the keyboard to move the cursor to the
desired entry field.
3.
Type the number(s) and press the Enter key on the keyboard to save
the entry and advance the cursor to the next entry field.
4.
Repeat steps 2 and 3 until all entries have been made.
5.
If desired, press [PRINT] to obtain a printout of the entered values.
6.
Press [TURN X-B ON] to enable the program if this key label is
displayed.