Device Description

The AMPLATZER™ TorqVue™ Delivery Sheath is designed to provide a pathway through which a device may be delivered. 
The body of each sheath is radiopaque for visibility under fluoroscopy. The included dilator eases penetration of tissue. The 
10 Fr, 12 Fr, and 13 Fr delivery sheaths also include a flush adapter to facilitate attachment of additional device components.

Indications for Use

The AMPLATZER™ TorqVue™ 45° x 45° Delivery Sheath is intended to provide a pathway through which devices are 
introduced within the chambers of the heart.

Contraindications

None known.

Warnings

• Do not use this device if the sterile package is open or damaged.
• To ensure component compatibility and correct fit when attaching any components other than an AMPLATZER™ loader to 

the 10 Fr, 12 Fr, or 13 Fr sheath, first attach the flush adapter to the sheath and then connect other components (including 
the hemostasis valve) to the flush adapter.

• Do not use a power injection syringe to inject contrast solution through this sheath.
• This device was sterilized with ethylene oxide and is for single use only. Do not reuse or resterilize this device. Attempts to 

resterilize this device can cause a malfunction, insufficient sterilization, or harm to the patient.

• Remove the dilator and sheath slowly to prevent an ingress of air.

Precautions

• Use on or before the last day of the expiration month that is printed on the product packaging label.
• This device should be used only by physicians who are trained in transcatheter techniques. The physician should 

determine which patients are suitable candidates for procedures using this device.

• The physician should exercise clinical judgment in situations that involve the use of anticoagulants or antiplatelet drugs 

before, during, and/or after the use of the TorqVue™ Delivery Sheath.

• Use caution when advancing the sheath and dilator to avoid damaging tissue and vessels or interfering with previously 

implanted medical devices.

• Prolonged procedures may result in increased exposure to anesthesia, contrast media, and/or radiation.
• Store in a dry place.

TorqVue™ 45° x 45° Delivery Sheath

Instructions for Use

Summary of Contents for AMPLATZER TorqVue Series

Page 1: ...sheath first attach the flush adapter to the sheath and then connect other components including the hemostasis valve to the flush adapter Do not use a power injection syringe to inject contrast solution through this sheath This device was sterilized with ethylene oxide and is for single use only Do not reuse or resterilize this device Attempts to resterilize this device can cause a malfunction ins...

Page 2: ...lock the components together 4 Place a 0 035 inch guidewire according the device s instructions for use 5 Advance the sheath and dilator over the guidewire 6 Turn the rotating luer on the dilator to unlock the components and slowly remove the dilator from the sheath WARNING Remove the dilator slowly to prevent an ingress of air 7 Remove the guidewire WARNING Connect a hemostasis valve to the sheat...

Page 3: ...PRESS OR IMPLIED INCLUDING ANY WARRANTY AS TO MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE See the Terms and Conditions of Sale for further information State of California USA Only WARNING This product can expose you to chemicals including ethylene oxide which is known to the State of California to cause cancer and birth defects or other reproductive harm For more information go to www P65W...

Page 4: ...r Outer diameter Length Usable length Recommended delivery sheath catheter dimensions Hydrophilic coating Federal law USA restricts this device to sale by or on the order of a physician or properly licensed practitioner Date of Manufacture Quantity Delivery Sheath ONLY Delivery Sheath ...

Reviews: